Performance Characteristics Appendix<br />

19.8 Performance Characteristics<br />

19.8.1 ECG Specifications<br />

Regulations<br />

Complies with the following regulations:<br />

ANSI/AAMI EC11 / EC13:2001/2002*<br />

ANSI/AAMI EC57:1998 (R2003)<br />

IEC 60601-2-27:2005 / EN60601-2-27:2006<br />

IEC 60601-2-25:1993 + A1:1999 / EN60601-2-25:1995 + A1:1999<br />

* Accuracy of input signal reproduction: Methods A & D were used, as prescribed in section 3.2.7.2 and 4.2.7.2 of<br />

this standard, to establish overall system error and frequency response. Because of the sampling characteristics<br />

and the asynchronism between sample rate and signal rate, the Passport V may produce a noticeable<br />

modulating effect from one cycle to the next, particularly in pediatric recordings. * Heart Rate Meter Accuracy<br />

and Response to Irregular Rhythm: Ventricular bigeminy: 80 bpm; Slow alternating ventricular bigeminy:<br />

60 bpm; Rapid alternating ventricular bigeminy: 120 bpm; Bidirectional systoles: 90 bpm. * Response<br />

time to HR change: From 80 to 120 bpm: less than 11 s; From 80 to 40 bpm: less than 11 s * Time to alarm for<br />

tachycardia: