ST Analysis (Optional) ECG Monitoring<br />

4.9 ST Analysis (Optional)<br />

NOTE: ST analysis is available for Adult and Pediatric patients only.<br />

P<br />

ISO Point<br />

Q<br />

R<br />

FIGURE 4-27 ST Monitoring<br />

ST deviation<br />

(Depression or Elevation)<br />

J Point<br />

S<br />

The formula for calculating ST segment offset value is: ST segment offset value = VST-VISO. If<br />

3-lead is detected, analysis provides one ST value from one ECG waveform (ECG1). If 5-lead<br />

is detected, analysis provides all seven ST values.<br />

The depression or elevation of the ST segment is measured as the vertical distance between<br />

the isoelectric (ISO) point which provides the baseline, and the ST point (See figure above).<br />

ST measurements are available on a maximum of three user selected ECG leads at a point<br />

situated 80 ms (heart rate 120 bpm or less) or 60 ms (heart rate more than 120 bpm) from<br />

the algorithmically determined end point of the QRS (J Point). In addition, the user can also<br />

select from three (3) different settings for the ST measurement point (80, 60, or 40 ms) from<br />

the J-point and independent of heart rate. These measurements are valid only on normal<br />

beats. Abnormal beats, like ventricular beats, are excluded from the analysis of the ST<br />

segment. Ventricular paced beats are also rejected from the analysis of the ST segment,<br />

because pacemaker tails distort the shape of the ST segment.<br />

ST segment changes are continuously measured by the monitor, but update of the displayed<br />

ST data is different depending on the ECG cable in use. When using a 3 or 5-lead ECG<br />

cable, the displayed ST data is updated approximately every 10 seconds.<br />

NOTE: The ST algorithm has been tested for accuracy of the ST<br />

segment data. The significance of the ST segment changes<br />

must be determined by a clinician.<br />

4 - 30 0070-10-0704-02 Passport V Operating Instructions<br />

ST Point<br />

T<br />

40 to 80 msec