Prompt Messages Messages<br />
CO 2 : Low Airway Press. Measurement of Flux<br />
Sensor in module is<br />
less than 400<br />
mmHg.<br />
CO 2: High Barometric Atmospheric pressure<br />
is too high(higher<br />
than 790mmHg).<br />
CO 2: Low Barometric Atmospheric pressure<br />
is too low(lower than<br />
428mmHg).<br />
CO 2: FilterLine Error Filterline is leaked or<br />
blocked.<br />
CO 2: Initialization Error No response after<br />
send order during<br />
initialization.<br />
CO 2: SelfTest Error Module can not work<br />
because of software<br />
fault,circuit error etc.<br />
CO 2 : Communication Error The monitor receives<br />
wrong responsion<br />
from CO 2 module.<br />
CO2: Communication Stop Sending out order is<br />
failed during work.<br />
CO2: Main Board Error Module has<br />
problems.<br />
CO 2: Replace Scrubber &<br />
Pump<br />
Module has<br />
problems.<br />
CO 2 : 15V Overrange Power supply to<br />
module is out of<br />
range.<br />
CO 2: Hardware Error Module has<br />
problems.<br />
CO 2 : No Watertrap Not water trap on<br />
CO 2 module<br />
Check airway's output<br />
and connection.<br />
If problems remain<br />
exist, return it to<br />
factory for<br />
maintaining.<br />
Check the CO 2<br />
connections,make sure<br />
that the monitor<br />
application site meets<br />
the requirements, and<br />
check for special<br />
sources that affect the<br />
ambient pressure.<br />
Restart the monitor.<br />
Check the CO 2<br />
connections,make sure<br />
that the monitor<br />
application site meets<br />
the requirements, and<br />
check for special<br />
sources that affect the<br />
ambient pressure.<br />
Restart the monitor.<br />
Check if there is a leak<br />
in the CO 2 sample line<br />
or the CO 2 sample line<br />
has been occluded.<br />
Contact Technical<br />
Support.<br />
Contact Technical<br />
Support.<br />
Contact Technical<br />
Support.<br />
Contact Technical<br />
Support.<br />
Reset the module.<br />
Contact Technical<br />
Support.<br />
Restart the monitor.<br />
Contact Technical<br />
Support.<br />
Check power supply.<br />
Restart the monitor.<br />
Contact Technical<br />
Support.<br />
Restart the monitor.<br />
Contact Technical<br />
Support.<br />
Make sure to plug-in<br />
water trap, and make<br />
sure it is firmly<br />
connected and<br />
fastened.<br />