Prompt Messages Messages<br />

CO 2 : Low Airway Press. Measurement of Flux<br />

Sensor in module is<br />

less than 400<br />

mmHg.<br />

CO 2: High Barometric Atmospheric pressure<br />

is too high(higher<br />

than 790mmHg).<br />

CO 2: Low Barometric Atmospheric pressure<br />

is too low(lower than<br />

428mmHg).<br />

CO 2: FilterLine Error Filterline is leaked or<br />

blocked.<br />

CO 2: Initialization Error No response after<br />

send order during<br />

initialization.<br />

CO 2: SelfTest Error Module can not work<br />

because of software<br />

fault,circuit error etc.<br />

CO 2 : Communication Error The monitor receives<br />

wrong responsion<br />

from CO 2 module.<br />

CO2: Communication Stop Sending out order is<br />

failed during work.<br />

CO2: Main Board Error Module has<br />

problems.<br />

CO 2: Replace Scrubber &<br />

Pump<br />

Module has<br />

problems.<br />

CO 2 : 15V Overrange Power supply to<br />

module is out of<br />

range.<br />

CO 2: Hardware Error Module has<br />

problems.<br />

CO 2 : No Watertrap Not water trap on<br />

CO 2 module<br />

Check airway's output<br />

and connection.<br />

If problems remain<br />

exist, return it to<br />

factory for<br />

maintaining.<br />

Check the CO 2<br />

connections,make sure<br />

that the monitor<br />

application site meets<br />

the requirements, and<br />

check for special<br />

sources that affect the<br />

ambient pressure.<br />

Restart the monitor.<br />

Check the CO 2<br />

connections,make sure<br />

that the monitor<br />

application site meets<br />

the requirements, and<br />

check for special<br />

sources that affect the<br />

ambient pressure.<br />

Restart the monitor.<br />

Check if there is a leak<br />

in the CO 2 sample line<br />

or the CO 2 sample line<br />

has been occluded.<br />

Contact Technical<br />

Support.<br />

Contact Technical<br />

Support.<br />

Contact Technical<br />

Support.<br />

Contact Technical<br />

Support.<br />

Reset the module.<br />

Contact Technical<br />

Support.<br />

Restart the monitor.<br />

Contact Technical<br />

Support.<br />

Check power supply.<br />

Restart the monitor.<br />

Contact Technical<br />

Support.<br />

Restart the monitor.<br />

Contact Technical<br />

Support.<br />

Make sure to plug-in<br />

water trap, and make<br />

sure it is firmly<br />

connected and<br />

fastened.<br />

14 - 14 0070-10-0704-02 Passport V Operating Instructions<br />

DPM<br />

DPM<br />

DPM<br />

DPM<br />

1DPM<br />

2Oridion<br />

1DPM<br />

2Oridion<br />

1DPM<br />

2Oridion<br />

1DPM<br />

2Oridion<br />

Oridion<br />

Oridion<br />

Oridion<br />

DPM<br />

DPM