Equipment Environment and Safety Specifications Appendix<br />

Transport<br />

Complies with the requirements of ISTA transport test procedure 2A (for shipping cargos in<br />

containers).<br />

Noise Level<br />

The patient monitor complies with the requirements of ISO 3744 in normal operation.<br />

Temperature controlled fan is used.<br />

Below and equal to 27 degree ambient temperature, noise ≤ 35 dBA measured at<br />

a distance of one meter from the device.<br />

Below and equal to 33 degree ambient temperature, noise ≤ 40 dBA measured at a<br />

distance of one meter from the device.<br />

Below and equal to 40 degree ambient temperature, noise ≤ 46 dBA measured at a<br />

distance of one meter from the device.<br />

Conditions: NIBP pump and valves not operational, local printer not operational, CO2 not operational.<br />

SPL range does not include patient heartbeat tone, key-pressing tone and alarm tone.<br />