Appendix Performance Characteristics<br />

No motion conditions 1 :<br />

PATIENT SIZE<br />

70% TO 100% 50% TO 69% 1-49%<br />

Adult Mode ± 2% ± 3% unspecified<br />

Pediatric Mode ± 2% ± 3% unspecified<br />

Neonatal Mode ± 3% ± 3% unspecified<br />

During motion conditions 2 :<br />

1 Masimo SET technology with LNOP and LNCS sensors have been validated for no motion accuracy in human<br />

blood studies on healthy adult volunteers in induced hypoxia studies in the range of 70-100% SpO2 against a<br />

laboratory co-oximeter and ECG monitor. This variation equals plus or minus one standard deviation, which<br />

encompasses 68% of the population.<br />

2 Masimo SET technology with LNOP and LNCS sensors have been validated for motion accuracy in human<br />

blood studies on healthy adult volunteers in induced hypoxia studies while performing rubbing and tapping<br />

motions, at 2 to 4 Hz at an amplitude of 1 to 2 cm and a non-repetitive motion between 1 to 5 Hz at an<br />

amplitude of 2 to 3 cm in induced hypoxia studies in the range of 70-100% SpO2 against a laboratory co-oximeter<br />

and ECG monitor. This variation equals plus or minus one standard deviation, which encompasses<br />

68% of the population.<br />

Resolution<br />

1%<br />

Average Time<br />

Optional 2-4 seconds, 4-6 seconds, 8 seconds, 10 seconds, 12 seconds, 14 seconds,<br />

16seconds<br />

Refresh Rate<br />

1s<br />

SpO 2 Response Time<br />

The response time is 18 seconds to 95% of final step change of % SpO2 value from 60 to<br />

95% at 75 bpm. Post averaging time is set at 8 seconds.<br />

Sensitivity<br />

SATURATION RANGE<br />

SATURATION RANGE<br />

PATIENT SIZE<br />

70% TO 100% 1-69%<br />

Adult Mode ± 3% unspecified<br />

Pediatric Mode ± 3% unspecified<br />

Neonatal Mode ± 3% unspecified<br />

Normal and high sensitivity modes are optional. The sensitivity mode affects the<br />

measurement of dim signals.<br />

Passport V Operating Instructions 0070-10-0704-02 19 - 17