Gas Menu Gas Monitoring (optional)<br />
Gas Menu<br />
11.3.1 Measure Unit<br />
MENU ITEM SELECTIONS COMMENTS<br />
Mixed Agent<br />
Audio<br />
On (default)<br />
Off<br />
The unit of measurement for CO2 can be manually selected in the Installation Menu. The<br />
choices are: mmHg, %, and kPa. The unit of measurement for O2 , N2O and Anesthetic<br />
Agent is %.<br />
11.3.2 Automatic identification of Anesthetic Agents<br />
Select to turn mixed agents audio alarm on or off.<br />
On: If Mixed agents are detected, the “Mixed<br />
Agent” alarm tone will sound.<br />
Off: If Mixed agents are detected, the “Mixed<br />
Agent” alarm tone will not sound.<br />
Pump Restart Select to restart the Gas Module pump.<br />
This selection is not available while the pump is<br />
functioning normally.<br />
The system shuts off the pump and gives a pump off<br />
alarm when a pneumatic error occurs.<br />
The pump can be restarted manually after the pump<br />
is shut off.<br />
Zero Gases Select to zero gas module.<br />
Zeroing is performed to eliminate the effects of<br />
baseline drift occurring during measurement.<br />
Manual zeroing is supported.<br />
A zeroing prompt should be provided during<br />
zeroing. A zeroing error alarm should be given in<br />
the case of zeroing failure.<br />
If Mixed Agents are identified, then the MAC value are invalidated and dashes are<br />
displayed. The actual value of primary agent concentration,N2O, CO2 and O2 data is<br />
displayed as usual.<br />
When the Anesthetic Agent is not identified the agent is labeled as "AGENT", no Anesthetic<br />
Agent related physiological alarms are activated.<br />
Menus and the corresponding display areas are refreshed once the Anesthetic Agent is<br />
identified.<br />
WARNING: When using the Gas Module, the maximum sampling rate at<br />
the nasal cannula is 200 ml/min (120 ml/min for Gas<br />
Module 3 with a neonatal water trap). This device should<br />
not be used on patients whose breathing could be impaired<br />
by this vacuum flow rate.<br />
NOTE: The Gas Module 3 is equipped with automatic barometric<br />
pressure compensation.<br />
11 - 6 0070-10-0704-02 Passport V Operating Instructions