Gas Menu Gas Monitoring (optional)<br />

Gas Menu<br />

11.3.1 Measure Unit<br />

MENU ITEM SELECTIONS COMMENTS<br />

Mixed Agent<br />

Audio<br />

On (default)<br />

Off<br />

The unit of measurement for CO2 can be manually selected in the Installation Menu. The<br />

choices are: mmHg, %, and kPa. The unit of measurement for O2 , N2O and Anesthetic<br />

Agent is %.<br />

11.3.2 Automatic identification of Anesthetic Agents<br />

Select to turn mixed agents audio alarm on or off.<br />

On: If Mixed agents are detected, the “Mixed<br />

Agent” alarm tone will sound.<br />

Off: If Mixed agents are detected, the “Mixed<br />

Agent” alarm tone will not sound.<br />

Pump Restart Select to restart the Gas Module pump.<br />

This selection is not available while the pump is<br />

functioning normally.<br />

The system shuts off the pump and gives a pump off<br />

alarm when a pneumatic error occurs.<br />

The pump can be restarted manually after the pump<br />

is shut off.<br />

Zero Gases Select to zero gas module.<br />

Zeroing is performed to eliminate the effects of<br />

baseline drift occurring during measurement.<br />

Manual zeroing is supported.<br />

A zeroing prompt should be provided during<br />

zeroing. A zeroing error alarm should be given in<br />

the case of zeroing failure.<br />

If Mixed Agents are identified, then the MAC value are invalidated and dashes are<br />

displayed. The actual value of primary agent concentration,N2O, CO2 and O2 data is<br />

displayed as usual.<br />

When the Anesthetic Agent is not identified the agent is labeled as "AGENT", no Anesthetic<br />

Agent related physiological alarms are activated.<br />

Menus and the corresponding display areas are refreshed once the Anesthetic Agent is<br />

identified.<br />

WARNING: When using the Gas Module, the maximum sampling rate at<br />

the nasal cannula is 200 ml/min (120 ml/min for Gas<br />

Module 3 with a neonatal water trap). This device should<br />

not be used on patients whose breathing could be impaired<br />

by this vacuum flow rate.<br />

NOTE: The Gas Module 3 is equipped with automatic barometric<br />

pressure compensation.<br />