NIBP Measurements NIBP Monitoring<br />

2. Try the calf. Irritable newborns will react to the cuff pressure but may tolerate the calf<br />

better than the arm. Place the cuff just above the ankle.<br />

3. Use the correct Newborn and Infant size cuffs. When applying, verify the cuff’s<br />

Index line falls between the Range lines.<br />

4. Try disposable cuffs. Disposable cuffs are more pliant than reusable ones. They<br />

generally fit smaller infants better.<br />

5. Place the cuff lightly. If the cuff is too snug, it won’t work properly. On infants, you should<br />

be able to easily move the cuff over the limb.<br />

7.4.16 NIBP List Tile<br />

NOTE: NIBP cannot be taken under all conditions. Even manual<br />

methods, employing a sphygmomanometer and<br />

stethoscope, will not work on unstable or active patients.<br />

The NIBP List tile can display up to five of the most recent NIBP measurements in row form.<br />

These measurements are displayed from newest (in the top row) to oldest (in the bottom row).<br />

As shown in the example of FIGURE 7-4, each row displays a time stamp, the three digit<br />

systolic and diastolic pressures separated with a “/”, and the three digit mean pressure in<br />

parentheses.<br />

FIGURE 7-4 Example NIBP List Tile<br />

The NIBP List tile can be configured from the Display Setup Menu to display in one of the<br />

numeric tiles (1 - 5) and it has a menu target labeled “NIBP” that opens the NIBP Menu.<br />

7.4.17 Adaptive Inflation Pressure<br />

During Interval or STAT mode, the inflation pressure is adjusted according to the previous<br />

reading of the systolic pressure. After the first successful measurement, the inflation pressure<br />

is the previous systolic +50 mmHg in the adult mode, +50 mmHg in the pediatric mode, and<br />

+40 mmHg in the neonate mode.<br />

If a manual measurement is taken between interval readings, the manual measurement uses<br />

Adaptive Inflation because the unit is in Interval Mode.<br />

7 - 12 0070-10-0704-02 Passport V Operating Instructions