Appendix Performance Characteristics<br />

Display Size<br />

The accuracy error above is less than ±5%; the change in display sensitivity complies with<br />

50.102.6 of IEC 60601-2-27:2005.<br />

1.25mm/mV (×0.125)<br />

2.5mm/mV (×0.25)<br />

5mm/mV (×0.5)<br />

10mm/mV (×1)<br />

20mm/mV (×2)<br />

40mm/mV (×4)<br />

ECG Signal Range<br />

+/- 8mV (peak-to-peak value); the output accuracy of signals complies with 50.102.1 of<br />

IEC60601-2-27:2005.<br />

Calibration Signal<br />

1mV (peak-to-peak value)<br />

Accuracy: 5% or 40µV, whichever is greater<br />

Detected RTI Range of Measured ECG Signals<br />

4.88 uV, ±8 mV when gain is ×0.125, ×0.25, ×0.5, ×1, ×2, ×4.<br />

ECG Calibration mode<br />

Manual Calibration<br />

Input Impedance<br />

> 5 Mohms<br />

Bandwidth (-3dB)<br />

Surgical mode: From 1 to 20 Hz<br />

Monitor mode: From 0.5 to 40 Hz<br />

ST Mode: From 0.05 to 40 Hz<br />

Diagnostic mode: From 0.05 to 150 Hz<br />

Passport V Operating Instructions 0070-10-0704-02 19 - 7