Respiration and CO 2 Troubleshooting Respiration Monitoring<br />

MESSAGE/PROBLEM REASON SOLUTION<br />

Resp Disturbed Patient Movement<br />

External Interference.<br />

Resp Rate Overrange Respiration Outside of<br />

measureable range<br />

Resp: Initialization Error During the Resp module poweron,as<br />

the Resp module<br />

communication stops,system<br />

fails to communicate with<br />

module.<br />

Resp: Communication Error System can't communicate<br />

correctly with Resp module,and<br />

receive the error data packets.<br />

Resp: Communication Stop As the Resp module<br />

communication stops, the data<br />

packets sent by the module<br />

can not be received.<br />

Resp: High Impedance Connections not tight and/or<br />

properly secured.<br />

Electrodes dry or loose.<br />

Cable or lead wires damaged.<br />

Check patient<br />

For impedance - Check<br />

Electrode Contacts / reposition<br />

electrodes/cable<br />

Check environment for source<br />

of interference.<br />

Check patient.<br />

Contact Technical Support.<br />

Restart the monitor.If the error<br />

still appears,please contact<br />

Technical Support.<br />

Contact Technical Support.<br />

Ensure proper connection.<br />

(Electrode to lead, lead to<br />

cable, cable to monitor).<br />

Re-prep skin and apply fresh,<br />

moist electrodes.<br />

Check with continuity tester<br />

5 - 6 0070-10-0704-02 Passport V Operating Instructions