ST Analysis (Optional) ECG Monitoring<br />

4.9.2.1 Adjusting the ISO and J/ST Point<br />

1. Open the ST Menu by using the Navigator Knob to select the ST parameter heading.<br />

ST Menu can also be accessed through the ECG Menu.<br />

2. Ensure the ST Analysis selection in the ST Menu is On. The Passport V will learn<br />

the patient's QRS complexes (one for a 3-lead leadset or 3 for a 5-lead leadset). These<br />

learned complexes will appear in the ST Menu with the monitor-selected ISO and J/ST<br />

points displayed.<br />

3. Scroll to the ISO selection to adjust the isoelectric point on the learned QRS complex.<br />

The isoelectric point is the area of the QRS complex following the P-wave and before the<br />

start of the Q-wave.<br />

4. Adjust the ISO point by pressing the Navigator Knob and turning to adjust the white<br />

ISO reference line. Press the Navigator Knob when the ISO point is satisfactory.<br />

5. Scroll to the J/ST point heading. Adjust the J/ST point by pressing the Navigator Knob<br />

and turning to adjust the orange and green J/ST reference lines. Press the Navigator<br />

Knob when the J/ST points are satisfactory.<br />

6. To adjust the ST-pt, scroll to the ST-pt heading and press the Navigator Knob to display<br />

a list of ST measurement point settings. Select the appropriate setting, then press the<br />

Navigator Knob when the ST-pt choice is satisfactory.<br />

4.9.2.2 Adjusting ST Measurement Points<br />

ST measurement points refer to ISO point (isoelectric reference point) and ST point. The<br />

Mortara algorithm uses the J+X method to position the ST point, which is located at X position<br />

on the right of J point.<br />

The following table describes the range of ST measurement points and their steps.<br />

POINT OPTION/RANGE STEP<br />

ISO -200 to -4ms 8ms<br />

J/ST 4 to 200ms 8ms<br />

ST-pt 40ms, 60ms, 80ms,<br />

60/80ms<br />

/<br />

ST-pt is set to 60 or 80ms from the J point. When HR is greater than 120 bpm, the distance<br />

between ST point and J point is set at 60ms; when HR is lower or equal to 120 bpm, the<br />

distance between ST point and J point is set at 80ms.<br />

4 - 32 0070-10-0704-02 Passport V Operating Instructions