Performance Characteristics Appendix<br />

Pace Pulse Rejection<br />

HR calculation rejects pulses with amplitude between ±2mV to ± 700 mV and width<br />

between 0.1 ms and 2 ms (no tails).<br />

To enable about 50% pace pulses with fig.5d (EC13, at page 28) waveforms to trigger<br />

pace pulse detector, the minimum input waveform change rate is 50 V/s RTI.<br />

ECG Specifications<br />

Mortara Arrhythmia ( V3.2 )<br />

HR range<br />

15 BPM to 300 BPM Adult,<br />

15 BPM to 350 BPM pediatric,<br />

15 BPM to 350 BPM neonatal<br />

Accuracy: '+/- 5 % for heart rates between 251 BPM' and 350 BPM, the greater of<br />

'+/- 3 BPM' or 3% for heart rates between 30 BPM and 250 BPM'<br />

Tall T waves Rejection<br />

Complies with 4.1.2.1 c of ANSI/AAMI EC13-2002.<br />

HR calculation is not affected for T-waves whose QRS complex is 100ms, QT interval is<br />

350ms, lasting time is 180ms and amplitude less than 1.2mv.<br />

ST Segment Analysis<br />

Scope: ST analysis is available for Adult and Pediatric patients only.<br />

ST Measurement Point<br />

ST measurement point can be manually adjusted: J+60 and J+80 (default: J+60 ms).<br />

ISO point can be manually adjusted: from 4 to 200 ms before R waves (default: 80ms)<br />

and adjusted interval: 4ms.<br />

J point can be manually adjusted: from 4 to 200 ms after R waves (default: 48ms).<br />

Real Time Reference of ST Segment Analysis: Supported.<br />

Refresh Rate of ST Reference: 16 beats.<br />

Arrhythmia Analysis<br />

Scope: Arrhythmia analysis is available for Adult and Pediatric patients only.<br />

Arrhythmia Type: 11 kinds of Arrhythmia events are included:<br />

ASYSTOLE, V-FIB, V-TACH, VRHYTHM, COUPLET, RUN, BIGEMINY, TRIGEMINY,<br />

IRREGULAR HR, BRADY, PVCs per minute<br />