Introduction Warnings<br />

Warnings<br />

WARNING: Use of the Passport V is restricted to one patient at a time.<br />

WARNING: This device is not intended for direct cardiac application.<br />

WARNING: Internal Electrical Shock Hazard - This unit does not contain any<br />

user-serviceable parts. Do not remove instrument covers. Refer<br />

Servicing to qualified personnel.<br />

WARNING: Do not use this monitor during MRI (Magnetic Resonance<br />

Imaging) scanning. Induced current could potentially cause<br />

burns. Accuracy of measurements on this unit and the MRI unit<br />

may also be affected.<br />

WARNING: To reduce the hazard of burns in the high-frequency surgical<br />

neutral electrode connection, the electrodes should not be<br />

located between the surgical site and the electro-surgical unit<br />

return electrode.<br />

WARNING: Ensure that conductive parts of the ECG electrodes do not<br />

contact other conductive parts including earth ground. Do not<br />

connect any non-isolated accessories to the Passport V or to the<br />

ECG or invasive pressure channel inputs when connected to a<br />

patient. Ensure that the total chassis leakage currents of all<br />

connected units does not exceed 300µA. Use an IEC 60601-1<br />

approved isolation / separation transformer if required. Do not<br />

simultaneously touch the patient and any piece of electrical<br />

equipment if any cover has been removed from the equipment.<br />

WARNING: Electrode polarization: some electrodes may be subject to large<br />

offset potentials due to polarization. Use only electrodes as<br />

recommended by <strong>Mindray</strong>. Recovery time after application of<br />

defibrillator pulses may be especially compromised. Squeeze<br />

bulb electrodes commonly used for diagnostic ECG recording<br />

may be particularly vulnerable to this effect. Electrodes of<br />

dissimilar metals should not be used unless the amplifier can<br />

handle polarization potentials as high as 1 volt (V).<br />

WARNING: The AC line cord and interface cables (i.e. non-patient cables)<br />

may utilize the same ground. Therefore, removal of the AC line<br />

cord does not necessarily isolate the Passport V if non-patient<br />

interface cables are attached.<br />

WARNING: Observe extreme caution when a defibrillator is used on a<br />

patient. Do not touch any part of patient, table, or monitor<br />

when a defibrillator is in use.<br />

WARNING: Do not incinerate battery; possible explosion may occur.<br />

viii 0070-10-0704-02 Passport V Operating Instructions