Prompt Messages Messages<br />

14.3 Prompt Messages<br />

14.3.1 ECG Prompt Message<br />

ALARM MESSAGE REASON SOLUTION<br />

ECG Self Test Error During Power cycle unit, if ECG<br />

module can not self test<br />

succussfully, this message will<br />

display.<br />

ECG: Communication Stop As the ECG module<br />

communication stops, the data<br />

packets sent by the module can<br />

not be received.<br />

ECG: Communication Error The ECG module communication<br />

error, the command can not be<br />

send correctly.<br />

14.3.2 Resp Prompt Message<br />

Power cycle unit. If message<br />

reappears, contact Technical<br />

Support.<br />

Contact Technical Support.<br />

Contact Technical Support.<br />

Leads Off Leads off patient. Connect lead wires.<br />

ESU-Resp Off The high frequency<br />

electrosurgery Unit<br />

Interference signal is detected.<br />

Wait that the high frequency<br />

electrosurgery Unit<br />

Interference disappears.<br />

ECG Signal Invalid Poor leads connection. Check Patient.<br />

Check Leads and leadlines.<br />

Pacer Rejection On When Pacer Reject is set to On. This is normal operation. When<br />

Pace Reject is set to Off, this<br />

message disappears.<br />

Learning Displayed when a learning cycle<br />

has been requested for<br />

Arrhythmia or ST.<br />

/<br />

No Arrhythmia Detection at<br />

Central<br />

Central Station does not have<br />

arrhythmia Analysis capability.<br />

ECG: Initialization Error During the ECG module poweron,as<br />

the ECG module<br />

communication stops,system fails<br />

to communicate with module.<br />

14 - 8 0070-10-0704-02 Passport V Operating Instructions<br />

/<br />

Contact Technical Support.<br />

ALARM MESSAGE REASON SOLUTION<br />

Resp: Initialization Error During the Resp module poweron,as<br />

the Resp module<br />

communication stops,system fails<br />

to communicate with module.<br />

Contact Technical Support.<br />

Resp: Communication Error System can't communicate Restart the monitor.If the error<br />

correctly with Resp module,and still appears,please contact<br />

receive the error data packets. Technical Support.