Prompt Messages Messages<br />
14.3 Prompt Messages<br />
14.3.1 ECG Prompt Message<br />
ALARM MESSAGE REASON SOLUTION<br />
ECG Self Test Error During Power cycle unit, if ECG<br />
module can not self test<br />
succussfully, this message will<br />
display.<br />
ECG: Communication Stop As the ECG module<br />
communication stops, the data<br />
packets sent by the module can<br />
not be received.<br />
ECG: Communication Error The ECG module communication<br />
error, the command can not be<br />
send correctly.<br />
14.3.2 Resp Prompt Message<br />
Power cycle unit. If message<br />
reappears, contact Technical<br />
Support.<br />
Contact Technical Support.<br />
Contact Technical Support.<br />
Leads Off Leads off patient. Connect lead wires.<br />
ESU-Resp Off The high frequency<br />
electrosurgery Unit<br />
Interference signal is detected.<br />
Wait that the high frequency<br />
electrosurgery Unit<br />
Interference disappears.<br />
ECG Signal Invalid Poor leads connection. Check Patient.<br />
Check Leads and leadlines.<br />
Pacer Rejection On When Pacer Reject is set to On. This is normal operation. When<br />
Pace Reject is set to Off, this<br />
message disappears.<br />
Learning Displayed when a learning cycle<br />
has been requested for<br />
Arrhythmia or ST.<br />
/<br />
No Arrhythmia Detection at<br />
Central<br />
Central Station does not have<br />
arrhythmia Analysis capability.<br />
ECG: Initialization Error During the ECG module poweron,as<br />
the ECG module<br />
communication stops,system fails<br />
to communicate with module.<br />
14 - 8 0070-10-0704-02 Passport V Operating Instructions<br />
/<br />
Contact Technical Support.<br />
ALARM MESSAGE REASON SOLUTION<br />
Resp: Initialization Error During the Resp module poweron,as<br />
the Resp module<br />
communication stops,system fails<br />
to communicate with module.<br />
Contact Technical Support.<br />
Resp: Communication Error System can't communicate Restart the monitor.If the error<br />
correctly with Resp module,and still appears,please contact<br />
receive the error data packets. Technical Support.