ECG Monitoring Relearning ST or Arrhythmia Analysis<br />

4.10 Relearning ST or Arrhythmia Analysis<br />

Automatic Relearning<br />

The Passport V initiates the learning process for ST measurements or Arrhythmia analysis<br />

after one of the following:<br />

Unit Power-Up<br />

Return to normal monitoring from Standby mode<br />

Enabling ST or Arrhythmia analysis<br />

The lead has been changed in ECG 1 waveform (3-lead only)<br />

Patient Size is changed<br />

Whenever the Relearn function is selected from the ST, ECG or Arrhythmia<br />

Menus<br />

The monitor switches between 3-lead and 5-lead mode<br />

Manual Relearning<br />

To initiate a manual relearn, select Relearn from the ECG Menu, Arrhythmia Menu, or<br />

ST Menu.<br />

It is recommended to initiate a manual Relearn after one or more of the following:<br />

The ECG electrodes have been repositioned<br />

Sufficient time has passed since the last Relearn<br />

After significant changes to the patient QRS complex<br />

After significant changes to the patient ECG rhythm<br />

A clinician has observed clinically questionable arrhythmia calls<br />

A Relearn must be initiated if “Learning” occurred during a “Leads Off” condition.<br />

Passport V Operating Instructions 0070-10-0704-02 4 - 33