Appendix Performance Characteristics<br />

after entering measurement mode, periodically for durations of typically 10 seconds at<br />

a rate of at most once per hour.<br />

If a change of 8 °C from the last AZ is detected.<br />

If a pressure change of 20 mmHg relative to the last ambient pressure measurement<br />

(less than the purge threshold) for a period of 30 seconds is detected. The module will<br />

be able to detect a real change in the ambient pressure and a pressure change due to<br />

partial blockage of the FilterLine.<br />

The module prevents the triggering of an SFM in the following situations:<br />

In case of purging until the end of this state.<br />

During a breath absence period which follows a valid breath.<br />

While waiting a minimum of 20 seconds for host SFM enable command. (After the 20second<br />

opportunity given to the host to schedule the SFM passes, the module<br />

schedules the SFM according to a priority determined by current conditions).<br />

System Response Time: The system response time (with a standard-length FilterLine)<br />

which includes the delay time and rise time (10% to 90%) in<br />

response to a step change in the CO2 concentration is 2.9<br />

seconds typical.<br />

Rise Time (Adult and<br />

Neonatal):<br />

DPM CO 2<br />

190 msec maximum<br />

Delay Time: 2.7 seconds typical<br />

Pump Calibration Interval: No routine calibration is required. The module should initially<br />

be calibrated after 1200 operating hours, then once a year<br />

or after 4,000 operating hours, whichever comes first.<br />

Initialization Time: Time to full accuracy (power on to first reading): Typically 30<br />

seconds for both reading and waveform (max 180 seconds).<br />

At full accuracy when value appears.<br />

CO2 CO2 measurement range: 0-99 mmHg<br />

Measurement accuracy: ±2mmHg (0-40mmHg)<br />

±5% of reading (41-76mmHg)<br />

±10% of reading (77-99mmHg)<br />

Resolution: 1mmHg<br />

Sampling rate: 50Hz<br />

Passport V Operating Instructions 0070-10-0704-02 19 - 23