Arrhythmia Alarms (optional) ECG Monitoring<br />

4.7.1.3 Ventricular Tachycardia (V-TACH) Alarm<br />

A V-TACH alarm is activated as follows:<br />

The range of the V-TACH rate is between 100 to 180 beats per minute.<br />

AND<br />

The V-TACH lethal arrhythmia alarm is activated when the configured number of<br />

consecutive PVCs is reached. The range for the V-TACH threshold is 3 – 15 beats per<br />

minute.<br />

A V-TACH alarm is a Priority 1 alarm event that produces:<br />

Alarm Priority 1 visual and audio alarm indicators.<br />

A V-TACH text message above the ECG1 waveform area.<br />

4.7.2 Non-Lethal Arrhythmia Alarms<br />

A Non-Lethal Arrhythmia is an arrhythmia that is most likely not life threatening to a patient.<br />

Bigeminy, Bradycardia, Couplet, Irregular Heart Rate, PVC/min, Run, Trigeminy, and<br />

Ventricular Rhythm (V-Rhythm) alarms are classified as non-lethal arrhythmia alarms. All other<br />

Non-Lethal Arrhythmias default to Alarm Priority 2. The Alarm Priority for non-lethal<br />

Arrhythmia is user-selectable between 1 and 2.<br />

NOTE: Non-lethal arrhythmia alarms are not latched alarms and<br />

can be acknowledged at any time. To acknowledge a nonlethal<br />

arrhythmia alarm, press the “Silence” key on the<br />

keypad.<br />

4.7.2.1 Bigeminy Alarm<br />

The Bigeminy alarm is activated when three or more cycles of one PVC coupled to one<br />

normal beat are detected.<br />

The Bigeminy alarm is an alarm event that produces:<br />

Alarm visual and audio alarm indicators.<br />

A BIGEMINY text message above the ECG1 waveform area.<br />

4.7.2.2 Brady (Bradycardia) Alarm<br />

The Brady alarm is activated when the heart rate falls to a value 10% lower than the user<br />

selected value for low heart rate alarm.<br />

The Brady alarm is an alarm event that produces:<br />

4.7.2.3 Couplet Alarm<br />

Alarm visual alarm indicator. Since Bradycardia is always accompanied by a low<br />

heart rate alarm, the Priority 1 audio alarm will sound.<br />

A Brady text message above the ECG1 waveform area.<br />

The Couplet alarm is activated when two consecutive PVCs are detected between normal<br />

beats.<br />

The Couplet alarm is an alarm event that produces:<br />

Alarm visual and audio alarm indicators.<br />

A COUPLET text message above the ECG1 waveform area.<br />

4 - 26 0070-10-0704-02 Passport V Operating Instructions