Passport® - Mindray
Passport® - Mindray
Passport® - Mindray
- TAGS
- mindray
- res.mindray.com
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
Arrhythmia Alarms (optional) ECG Monitoring<br />
4.7.1.3 Ventricular Tachycardia (V-TACH) Alarm<br />
A V-TACH alarm is activated as follows:<br />
The range of the V-TACH rate is between 100 to 180 beats per minute.<br />
AND<br />
The V-TACH lethal arrhythmia alarm is activated when the configured number of<br />
consecutive PVCs is reached. The range for the V-TACH threshold is 3 – 15 beats per<br />
minute.<br />
A V-TACH alarm is a Priority 1 alarm event that produces:<br />
Alarm Priority 1 visual and audio alarm indicators.<br />
A V-TACH text message above the ECG1 waveform area.<br />
4.7.2 Non-Lethal Arrhythmia Alarms<br />
A Non-Lethal Arrhythmia is an arrhythmia that is most likely not life threatening to a patient.<br />
Bigeminy, Bradycardia, Couplet, Irregular Heart Rate, PVC/min, Run, Trigeminy, and<br />
Ventricular Rhythm (V-Rhythm) alarms are classified as non-lethal arrhythmia alarms. All other<br />
Non-Lethal Arrhythmias default to Alarm Priority 2. The Alarm Priority for non-lethal<br />
Arrhythmia is user-selectable between 1 and 2.<br />
NOTE: Non-lethal arrhythmia alarms are not latched alarms and<br />
can be acknowledged at any time. To acknowledge a nonlethal<br />
arrhythmia alarm, press the “Silence” key on the<br />
keypad.<br />
4.7.2.1 Bigeminy Alarm<br />
The Bigeminy alarm is activated when three or more cycles of one PVC coupled to one<br />
normal beat are detected.<br />
The Bigeminy alarm is an alarm event that produces:<br />
Alarm visual and audio alarm indicators.<br />
A BIGEMINY text message above the ECG1 waveform area.<br />
4.7.2.2 Brady (Bradycardia) Alarm<br />
The Brady alarm is activated when the heart rate falls to a value 10% lower than the user<br />
selected value for low heart rate alarm.<br />
The Brady alarm is an alarm event that produces:<br />
4.7.2.3 Couplet Alarm<br />
Alarm visual alarm indicator. Since Bradycardia is always accompanied by a low<br />
heart rate alarm, the Priority 1 audio alarm will sound.<br />
A Brady text message above the ECG1 waveform area.<br />
The Couplet alarm is activated when two consecutive PVCs are detected between normal<br />
beats.<br />
The Couplet alarm is an alarm event that produces:<br />
Alarm visual and audio alarm indicators.<br />
A COUPLET text message above the ECG1 waveform area.<br />
4 - 26 0070-10-0704-02 Passport V Operating Instructions