Performance Characteristics Appendix<br />

NIBP Start Pressure Settings and Ranges<br />

The Start Pressure is adjustable throughout the following ranges and is set to the following<br />

defaults:<br />

USER-SELECTABLE START<br />

DEFAULT<br />

PATIENT SIZE PRESSURE RANGE INCREMENT SIZE SETTING<br />

Adult Mode 100-280 mmHg 5 mmHg 180 mmHg<br />

Pediatric Mode 60-180 mmHg 5 mmHg 140 mmHg<br />

Neonatal Mode 40-120 mmHg 5 mmHg 100 mmHg<br />

NIBP Measurement Cycle<br />

The Maximum Measurement Cycle Duration is 180 sec for Adult and Pediatric patients<br />

(Reference: EN60601-2-30, 1995).<br />

The Maximum Measurement Cycle Duration is 90 sec for Neonatal patients (Reference:<br />

EN60601-2-30, 1995).<br />

Pulse Rate Range for NIBP Measurement<br />

PATIENT SIZE RANGE ACCURACY<br />

Adult Mode 35 - 245 bpm The greater of ± 3 bpm or ± 3%<br />

Pediatric Mode 35 - 245 bpm The greater of ± 3 bpm or ± 3%<br />

Neonatal Mode 70 - 245 bpm The greater of ± 3 bpm or ± 3%<br />

19.8.5 Masimo SET SpO 2 Performance Requirements<br />

Sensor Compatibility<br />

LNOP ® and LNCS ® Series<br />

Measured Parameters<br />

SpO 2 and PR<br />

SpO 2 Measurement Range<br />

From 1% to 100%<br />

SpO 2 Accuracy<br />

A functional tester or patient simulator not used to assess the accuracy of the pulse oximeter<br />

probe or a pulse oximeter monitor.<br />

19 - 16 0070-10-0704-02 Passport V Operating Instructions