CO 2 Menu CO 2 Monitoring (optional)<br />

NOTE: The watertrap has a filter to prevent bacterium, water, and<br />

secretions from entering the module. After long-term use,<br />

dust or other substances may compromise the performance<br />

of the filter or even block the airway. In this case, replace<br />

the watertrap. It is recommended to replace the watertrap<br />

once every month, or when the watertrap is found leaky,<br />

damaged, or contaminated.<br />

FIGURE 10-5 CO 2 Setup Menu (DPM Sidestream Only)<br />

CO 2 Setup Menu (DPM Sidestream Only)<br />

MENU ITEM SELECTIONS COMMENTS<br />

Previous Screen — Select to return to the previous menu.<br />

Humidity On<br />

Select On for Body Temperature Pressure<br />

Compensation Off (default)<br />

Saturation (BTPS).<br />

Select Off for Ambient Temperature and Pressure<br />

Dry (ATPD).<br />

O 2<br />

Compensation<br />

Note: If using the Oridion Medical<br />

miniMediCO 2 module, the humidity<br />

compensation is always on.<br />

0 (default) to 100% Select the O 2 level used for O 2 compensation.<br />

Note: The sum of compensation values input for<br />

oxygen, nitrous oxide and desflurane must not<br />

be greater than 100%.<br />

10 - 6 0070-10-0704-02 Passport V Operating Instructions