30.01.2013 Views

Passport® - Mindray

Passport® - Mindray

Passport® - Mindray

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

Quick Trends Trends<br />

15.2 Quick Trends<br />

The Quick Trend display allows the user to view an abbreviated numeric listing of HR,<br />

SpO2 and NIBP data only. To access this display from the normal monitoring screen, press<br />

the TRENDS key once. This display may also be accessed from the other trend displays via<br />

a menu choice. A maximum of 9 time-stamped entries may be stored. When the maximum<br />

number of entries has been reached, the oldest entry will be deleted from the trend record in<br />

order to allow storage of a new entry.<br />

FIGURE 15-1 Quick Trends<br />

Clearing Trend Data<br />

To manually clear all trend data, including Graphic Trends, choose Clear Trends from the<br />

menu. A confirmation prompt will appear. Once cleared, the data cannot be restored.<br />

All trend data is automatically cleared when the patient is discharged from the monitor.<br />

All trend data is also cleared if the monitor’s displayed time or date is changed.<br />

Removing the Quick Trend Display<br />

The Quick Trend display does not automatically time-out and must be manually removed to<br />

return to the normal waveform display. To remove the Quick Trend display, choose<br />

Normal Screen from the menu, or press the NORMAL SCREEN key.<br />

15 - 2 0070-10-0704-02 Passport V Operating Instructions