Performance Characteristics Appendix<br />

19.8.2 RESP<br />

Function and Specification<br />

Measured Parameters: RR, Waveform<br />

Measurement Method: Trans-thoracic impedance<br />

Intended Patient: Adult, Pediatric, Neonatal<br />

Measured Lead: Optional RA-LA or RA-LL; default RA-LL<br />

Waveform Gain Adjustment: Optional 1, 2, 3, 4, and 5<br />

Waveform Refresh Velocity: Optional 3.125, 6.25, 12.5 and 25.0mm/s<br />

Differential Input Impedance: More than 5Mohms<br />

Excitation Current: < 620 µA RMS<br />

Baseline Impedance Range: From 200 to 2500 ohms (use the ECG cable with 1 kohms<br />

resistance)<br />

Respiration Impedance Range: From 0.3 to 5 ohms<br />

Linear Signal Range: Minimum 3 ohms p-p<br />

Bandwidth: From 0.2 to 2 Hz (-3 dB)<br />

RTI Noise: < 0.1 ohms (peak value) (with standard ECG cables and patient impedance<br />

500 ohms)<br />

RR spec<br />

Range<br />

Adult: From 0 to 120 RPM<br />

Pediatric and neonatal: From 0 to 150 RPM<br />

Resolution: 1 RPM<br />

Accuracy: From 6 to 150 RPM: ±2 RPM or ±2%, whichever is greater<br />

From 0 to 5 RPM: not specified<br />

Apnea Alarm<br />

Artifact (CVA) Notification<br />

Screen Display and UI Element<br />

RESP Normal Display<br />

Waveform Display<br />

Respiration waveform (thoracic impedance)<br />

Waveform label<br />

Waveform gain<br />

Lead selected<br />

RR Source<br />

Data Storage<br />

Respiration Rate<br />

19 - 12 0070-10-0704-02 Passport V Operating Instructions