System Configuration Installation Menu<br />

MENU TITLE<br />

ON SCREEN MENU CHOICES DEFAULT COMMENTS<br />

Notch Filter 50 Hz<br />

60 Hz<br />

Off<br />

ESU Filter Auto<br />

Disable<br />

SpO 2 Sensor Off<br />

Audio<br />

Off<br />

Once<br />

Repeat<br />

SpO 2 Tones Standard Tones<br />

Alternate Tones<br />

Enable Arrhythmia All<br />

Off Selection<br />

Yes<br />

No<br />

Apnea Latch On<br />

Off<br />

Enable Silence All<br />

Permanent Selection<br />

Yes<br />

No<br />

Auxiliary Output Nurse call<br />

Analog Output<br />

Defib Sync<br />

Nurse Call Mode Normally Open<br />

Normally Close<br />

Set up Serial Port 1 None<br />

DIAP<br />

Gas Module<br />

Set up Serial Port 2 None<br />

DIAP<br />

Gas Module<br />

Copy Monitor Defaults<br />

to Storage Device<br />

Copy Monitor Defaults<br />

from Storage Device<br />

Re-boot in demo mode No<br />

Yes<br />

Restore factory<br />

defaults<br />

Advanced Installation<br />

Setup<br />

60 Hz Select to change notch filter<br />

for ECG.<br />

Auto Select to change the ESU<br />

filter.<br />

Off Select to change the SpO 2<br />

Sensor Off Audio alert<br />

type.<br />

Standard Tones Select to change the SpO 2<br />

tones.<br />

No Select to enable or disable<br />

the Arrhythmia All Off<br />

menu selection.<br />

On Select to turn apnea alarm<br />

latching on or off.<br />

No Select to enable or disable<br />

the Permanent Audio Off<br />

menu selection.<br />

Nurse Call Select to change the<br />

function of the Auxiliary<br />

Output interface.<br />

Normally Open Select to change the type<br />

of Nurse Call system<br />

None Select to set up comm port.<br />

None Select to set up comm port.<br />

Select to copy monitor<br />

defaults to Storage Device.<br />

Select to copy monitor<br />

defaults from Storage<br />

Device.<br />

No Set to “YES” to start the<br />

monitor in demonstration<br />

mode on next power-up.<br />

Normal monitoring will<br />

resume after cycling power<br />

in demonstration mode.<br />

Select to restore factory<br />

defaults<br />

Select to access Advanced<br />

Installation Setup Menu.<br />

Change Password Select to change<br />

password.<br />

Options Select to add/view<br />

options.<br />

Passport V Operating Instructions 0070-10-0704-02 2 - 3