Trends OxyCRG<br />

15.5 OxyCRG<br />

The OxyCRG function is available for neonatal patients only.<br />

The OxyCRG display allows the user to view a graphic summary of 5 specific patient<br />

parameters, including HR, SpO2, Resp Wave, Temp and IBP1.<br />

To access this display from the normal screen, press the TRENDS key four times (Patient Size<br />

must be set to Neonate to view OxyCRG). This display may also be accessed from the other<br />

trend displays via a menu choice.<br />

FIGURE 15-4 OxyCRG Display<br />

The events that trigger the storage of OxyCRG data are Interval events, Alarm events, and<br />

Manually marked events. The interval for triggering the interval event is 2s maximally.<br />

The parameters stored for the OxyCRG are instantaneous values of HR, SpO2, RR, Resp<br />

Wave, Temp, and IBP1. The time marker is one second. Trended data includes the Time<br />

marker, Parameter value, and Event type.<br />

OxyCRG screen displays the trend graph, parameter values, and events. The rules for<br />

displaying trend graph are as follows:<br />