7 Optimizing Your Method<br />
Modifying an Automatic Tune<br />
• Tune Resolution—Select this option to tune the resolution by itself. This resolution<br />
will tune the system at 100 and 1000 amu/s scan rates. You may also tune at a higher<br />
scan rate.<br />
• Mass—Use this pull-down menu to select the ion to be used for tuning.<br />
• Peak Width—Use this field to enter the target peak width.<br />
• Measure at %—Use this pull-down menu to select the location on the peak at which<br />
you want to measure the target peak width.<br />
– 10—Select this option to measure the width at 10% <strong>of</strong> the peak height.<br />
– 50—Select this option to measure the width at 50% <strong>of</strong> the peak height.<br />
• Lens Tune Relation—Use this pull-down menu to set the occurrence <strong>of</strong> the<br />
resolution tune parameters before or after a lens tune.<br />
– Before and After—Select this option to use the same resolution parameters<br />
before and after a lens tune.<br />
– Before—Select this option to use the resolution parameters before a lens tune.<br />
– After—Select this option to use the resolution parameters after a lens tune.<br />
• Tune High Scan Rate—Select this checkbox to tune the resolution at a scan rate in<br />
addition to the 100 and 1000 amu/s default scan rates.<br />
Thermo Scientific <strong>ISQ</strong> <strong>User</strong> <strong>Guide</strong> 85