7 Optimizing Your Method<br />

Modifying an Automatic Tune<br />

• Tune Resolution—Select this option to tune the resolution by itself. This resolution<br />

will tune the system at 100 and 1000 amu/s scan rates. You may also tune at a higher<br />

scan rate.<br />

• Mass—Use this pull-down menu to select the ion to be used for tuning.<br />

• Peak Width—Use this field to enter the target peak width.<br />

• Measure at %—Use this pull-down menu to select the location on the peak at which<br />

you want to measure the target peak width.<br />

– 10—Select this option to measure the width at 10% <strong>of</strong> the peak height.<br />

– 50—Select this option to measure the width at 50% <strong>of</strong> the peak height.<br />

• Lens Tune Relation—Use this pull-down menu to set the occurrence <strong>of</strong> the<br />

resolution tune parameters before or after a lens tune.<br />

– Before and After—Select this option to use the same resolution parameters<br />

before and after a lens tune.<br />

– Before—Select this option to use the resolution parameters before a lens tune.<br />

– After—Select this option to use the resolution parameters after a lens tune.<br />

• Tune High Scan Rate—Select this checkbox to tune the resolution at a scan rate in<br />

addition to the 100 and 1000 amu/s default scan rates.<br />