14.12.2012 Views

ISQ User Guide - Write Frame of Mind

ISQ User Guide - Write Frame of Mind

ISQ User Guide - Write Frame of Mind

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

3 Tuning the <strong>ISQ</strong><br />

Using <strong>ISQ</strong> Autotune<br />

• EI Full Tune—This tune is used to completely retune the system. It takes the longest<br />

amount <strong>of</strong> time to run, but the advantage <strong>of</strong> using it is that it re-optimizes nearly all the<br />

parameters affecting the signal. This type <strong>of</strong> tune will perform a mass calibration, tune the<br />

lenses and resolution, and perform a leak check. The detector gain will be calibrated to<br />

generate 300,000 electrons for every ion that strikes the detector. You should run an EI<br />

full tune when the daily tune or the daily tune check is not adequate, when the electron<br />

multiplier is getting old (tuning to high electron multiplier voltages), or the first tune<br />

after you replace the electron multiplier. Unless your SOP requires it, this is not the best<br />

tune to use on a daily basis because <strong>of</strong> the length <strong>of</strong> time it takes to run it.<br />

• CI- Tune—This tune is used to analyze samples with negative CI. The standard NCI<br />

tune performs a mass calibration, then tunes the lenses and sets the resolution. This type<br />

<strong>of</strong> tune assumes you are using methane as the CI reagent gas and tunes the system with a<br />

1.0 mL/min flow. This tune does not set the detector gain.<br />

Note Chemical ionization tunes are very different from the electron ionization tunes. You<br />

should not use a CI tune unless your instrument has a CI ion volume and CI reagent gas<br />

installed.<br />

• CI+ Tune—This tune is used to analyze samples with positive CI. The standard PCI tune<br />

performs a mass calibration, then tunes the lenses and sets the resolution. This type <strong>of</strong><br />

tune assumes you are using methane as the CI reagent gas and tunes the system with a<br />

1.5 mL/min flow. This tune does not set the detector gain.<br />

Note To add a tune type to the list, see Modifying an Automatic Tune.<br />

3. Select the Display Report When Complete checkbox so that you can view the tune<br />

report after running the tune.<br />

4. Select the Show Spectra checkbox to show the spectra while the system is tuning.<br />

Thermo Scientific <strong>ISQ</strong> <strong>User</strong> <strong>Guide</strong> 19

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!