8 Troubleshooting<br />

Reconfiguring Your Instrument<br />

12. Select the Filament 2 checkbox to enable the maintenance reminder. Then set the<br />

number <strong>of</strong> days in which you want to be reminded to check filament 2. In a leak-free<br />

system, the filament should last between 30-360 days, depending on usage<br />

13. Select the Ion source checkbox to enable the maintenance reminder. Then set the<br />

number <strong>of</strong> days in which you want to be reminded to check the ion source. The time<br />

between cleaning depends very strongly on your analysis. You will have to determine the<br />

correct length <strong>of</strong> time between source cleanings.<br />

Thermo Scientific <strong>ISQ</strong> <strong>User</strong> <strong>Guide</strong> 103