14.12.2012 Views

ISQ User Guide - Write Frame of Mind

ISQ User Guide - Write Frame of Mind

ISQ User Guide - Write Frame of Mind

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

4 Creating a Method<br />

b. If you have a split / splitless inlet (SSL) injector, click the SSL tab to configure the<br />

injector port settings.<br />

The inlet should be turned on and set to a temperature that is at least 10 °C<br />

higher than the boiling point <strong>of</strong> your least volatile analyte. The material injected<br />

into the port will vaporize and move into the GC column quickly. Higher<br />

temperatures can lead to thermal decomposition <strong>of</strong> some analytes, so you will<br />

have to optimize the injector temperature for your analysis. Set the SSL<br />

temperature between 50 and 375 °C (a typical value would be 225 °C).<br />

Useful for diluting high concentrations <strong>of</strong> sample, the split flow is the amount <strong>of</strong><br />

gas that is swept through the injector to the exhaust port. Higher values will give<br />

more dilution. The split flow will also reduce the amount <strong>of</strong> contamination that<br />

builds up in your system. The split flow ratio is the ratio <strong>of</strong> the split flow to the<br />

carrier gas flow. It is effectively the dilution ratio <strong>of</strong> the sample. This is typically<br />

turned on and set to a flow <strong>of</strong> 50 mL/min. However, this increases the carrier gas<br />

usage, so for your analysis, lower split flows may be more acceptable.<br />

We recommend turning on the septum purge, which means an additional 5<br />

mL/min <strong>of</strong> carrier gas will go through the injector. This reduces the buildup <strong>of</strong><br />

contaminants in the injector, on the column, and in the <strong>ISQ</strong>. If you perform a<br />

splitless injection, even if the split flow is set, the split flow will be turned <strong>of</strong>f for<br />

the splitless time. The septum purge will be turned <strong>of</strong>f for the stop purge time.<br />

After these times, the split flow and septum purge will be reactivated.<br />

Finally, if your analysis requires a higher flow to quickly sweep the analytes into<br />

the column, which may be needed with high temperature injectors and thermally<br />

labile compounds, you can use the surge pressure to increase the column flow for<br />

the surge duration time.<br />

46 <strong>ISQ</strong> <strong>User</strong> <strong>Guide</strong> Thermo Scientific

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!