Investigating Baseline Issues<br />

Table 2. Troubleshooting Baseline Issues in Your Data<br />

Behavior Cause Characteristic Remedy<br />

Drifting<br />

Irregular Shape<br />

General<br />

Falling<br />

Rising<br />

Rising baseline when<br />

oven temperature is high<br />

Peak areas are not linear<br />

with concentration<br />

Chromatographic peak is<br />

wrong shape<br />

S-shaped<br />

8 Troubleshooting<br />

Investigating Baseline Issues<br />

Stationary phase has<br />

accumulated in column<br />

Replace the column or cut <strong>of</strong>f the end <strong>of</strong> the column.<br />

Chromatographic baseline is high Replace the column or cut <strong>of</strong>f the end <strong>of</strong> the column.<br />

Carrier gas pressure is too low<br />

Carrier gas flow is drifting<br />

Impurities have accumulated in<br />

column<br />

Replace the carrier gas cylinder or increase the<br />

pressure. Make sure the vacuum compensation is on.<br />

Check the carrier gas tank and Digital Carrier<br />

Controller (DCC).<br />

Check impurity levels in your gas. Use the correct gas<br />

purity and filter. Check your septum and replace, if<br />

necessary.<br />

Carrier gas leak in the system<br />

Perform a leak test and tighten the connections to the<br />

carrier gas line.<br />

Column is baking out Wait for the column to stabilize.<br />

Impurities have accumulated in<br />

column<br />

Check impurity levels in the gas source. Use correct<br />

gas purity.<br />

Column is contaminated Recondition or replace the column.<br />

Surface activity or poor sample<br />

preparation<br />

GC column has oxygen or is<br />

damaged, injector port problems<br />

Excessive column bleed at high<br />

oven temperatures<br />

Oxygen contamination is<br />

decomposing the stationary phase<br />

Verify your standards and sample preparation<br />

method.<br />

Trim or change the column.<br />

Change the liner.<br />

Change the column stationary phase.<br />

Reduce the column temperature.<br />

Bake out the column.<br />

Install a high-temperature column.<br />

Install oxygen filters in carrier gas line.<br />

Check pneumatic and inlet systems for leaks.<br />

Use correct gas purity with low oxygen content.<br />

Thermo Scientific <strong>ISQ</strong> <strong>User</strong> <strong>Guide</strong> 91