Investigating Baseline Issues<br />
Table 2. Troubleshooting Baseline Issues in Your Data<br />
Behavior Cause Characteristic Remedy<br />
Drifting<br />
Irregular Shape<br />
General<br />
Falling<br />
Rising<br />
Rising baseline when<br />
oven temperature is high<br />
Peak areas are not linear<br />
with concentration<br />
Chromatographic peak is<br />
wrong shape<br />
S-shaped<br />
8 Troubleshooting<br />
Investigating Baseline Issues<br />
Stationary phase has<br />
accumulated in column<br />
Replace the column or cut <strong>of</strong>f the end <strong>of</strong> the column.<br />
Chromatographic baseline is high Replace the column or cut <strong>of</strong>f the end <strong>of</strong> the column.<br />
Carrier gas pressure is too low<br />
Carrier gas flow is drifting<br />
Impurities have accumulated in<br />
column<br />
Replace the carrier gas cylinder or increase the<br />
pressure. Make sure the vacuum compensation is on.<br />
Check the carrier gas tank and Digital Carrier<br />
Controller (DCC).<br />
Check impurity levels in your gas. Use the correct gas<br />
purity and filter. Check your septum and replace, if<br />
necessary.<br />
Carrier gas leak in the system<br />
Perform a leak test and tighten the connections to the<br />
carrier gas line.<br />
Column is baking out Wait for the column to stabilize.<br />
Impurities have accumulated in<br />
column<br />
Check impurity levels in the gas source. Use correct<br />
gas purity.<br />
Column is contaminated Recondition or replace the column.<br />
Surface activity or poor sample<br />
preparation<br />
GC column has oxygen or is<br />
damaged, injector port problems<br />
Excessive column bleed at high<br />
oven temperatures<br />
Oxygen contamination is<br />
decomposing the stationary phase<br />
Verify your standards and sample preparation<br />
method.<br />
Trim or change the column.<br />
Change the liner.<br />
Change the column stationary phase.<br />
Reduce the column temperature.<br />
Bake out the column.<br />
Install a high-temperature column.<br />
Install oxygen filters in carrier gas line.<br />
Check pneumatic and inlet systems for leaks.<br />
Use correct gas purity with low oxygen content.<br />
Thermo Scientific <strong>ISQ</strong> <strong>User</strong> <strong>Guide</strong> 91