4 Creating a Method<br />

To use the hot needle technique, select the standard injection depth and use a<br />

pre-injection dwell time <strong>of</strong> at least 2 seconds. To use the cold needle technique,<br />

select a minimum injection depth and do not use a pre-injection dwell time. The<br />

post injection dwell time can be used to bake out the needle and ensure that all<br />

semi-volatile compounds are cleaned out <strong>of</strong> the system before the next injection<br />

is made.<br />

4. Click the <strong>ISQ</strong> icon in the left pane to create a method for the <strong>ISQ</strong>.<br />

a. From the Method Type pull-down menu, select Acquisition. An Acquisition method<br />

is used to collect data.<br />

Note The maintenance method is used to bake out or cool down the <strong>ISQ</strong> during a<br />

sequence. This can be useful if you know you want to perform these tasks and you want to<br />

do them in an automated way as part <strong>of</strong> your data acquisition.<br />

34 <strong>ISQ</strong> <strong>User</strong> <strong>Guide</strong> Thermo Scientific