14.12.2012 Views

ISQ User Guide - Write Frame of Mind

ISQ User Guide - Write Frame of Mind

ISQ User Guide - Write Frame of Mind

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

. Configure the Pre-Injection group:<br />

4 Creating a Method<br />

• Pre-Injection—Use the Solvent and Cycles fields to set the number <strong>of</strong> solvent<br />

purges that will occur before the autosampler touches your sample. You should<br />

always have some sample rinses, either before or after injection, to make sure you<br />

do not have sample carryover from one injection to the next. You can configure<br />

the settings so that the syringe is purged with the same solvent that was used in<br />

location A, B, C or D or with solvents A and B or C and D. Typically, there is 0<br />

or 1 cycles <strong>of</strong> pre-injection purges.<br />

• Sample—Use the Rinses pull-down menu to select the number <strong>of</strong> times the<br />

syringe is rinsed with your sample before each injection. Rinses help ensure the<br />

sample being injected is not diluted by the residual rinse solvents. By purging the<br />

syringe with your sample before injection, the dilution is minimal. The standard<br />

setting is between 1 and 3 rinses. If you have a very limited amount <strong>of</strong> sample,<br />

however, you may want to set this field to 0 to conserve the sample.<br />

• Post-Injection—Use the Solvent and Cycles fields to set the number <strong>of</strong> solvent<br />

purges that will occur after the autosampler touches your sample. You should<br />

always have some sample rinses, either before or after injection, to make sure you<br />

do not have sample carryover from one injection to the next. You can have the<br />

syringe purged with the same solvent that was used in location A, B, C or D or<br />

with solvents A and B or C and D. Typically, there are 1 to 5 cycles <strong>of</strong><br />

post-injection purges.<br />

c. Configure the Injection group:<br />

• Injection Depth—Use the pull-down menu to select Standard or Minimum to<br />

indicate how the sample is introduced into the GC. If you select Standard, the<br />

autosampler will insert the needle all the way into the injection port. If you select<br />

Minimum, the autosampler will barely enter the injection port.<br />

• Pre-inj Dwell Time(s)—Use this field to enter the time (in seconds) that the<br />

needle will be in the injection port before the plunger injects the sample.<br />

• Post-inj Dwell Time (s)—Use this field to enter the time (in seconds) that the<br />

needle will be in the injection port after the plunger injects the sample.<br />

The two main forms <strong>of</strong> injection are a hot needle technique and a cold needle<br />

technique.<br />

In a hot needle injection, the needle is preheated before the sample is introduced,<br />

which causes the sample to vaporize before it leaves the needle. The vapor is then<br />

mixed with carrier gas and swept into the column.<br />

In a cold-needle injection, the needle is kept as cold as possible so that the sample<br />

can be introduced as a liquid stream that vaporizes in the injector. The sample<br />

vapor is produced closer to the column inlet, which can lead to better sensitivity,<br />

but worse reproducibility from one run to the next.<br />

Thermo Scientific <strong>ISQ</strong> <strong>User</strong> <strong>Guide</strong> 33

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!