8 Troubleshooting<br />

Reconfiguring Your Instrument<br />

7. Click the Communications button to reset the network IP address and assign a TCP<br />

port. If you have other equipment on your network and you need to move the <strong>ISQ</strong> to a<br />

different IP address, you can enter it here. For security purposes, you may also want to<br />

modify the TCP port. Consult your local IT Department for help.<br />

8. Click the Maintenance Intervals tab to set the number <strong>of</strong> days until you plan to perform<br />

maintenance on certain components <strong>of</strong> your GC/MS system. As a default, the foreline<br />

pump and turbo pump oil are automatically enabled. You can monitor the progress <strong>of</strong><br />

these settings in the Status pane <strong>of</strong> Xcalibur.<br />

9. Select the Foreline pump oil checkbox to enable the maintenance intervals. Then set the<br />

number <strong>of</strong> days in which you want to be reminded to check the oil. The manufacturer<br />

recommends changing the oil every 125 days. If you are using corrosive gases, such as<br />

ammonia, you should change the oil every 30 days.<br />

Thermo Scientific <strong>ISQ</strong> <strong>User</strong> <strong>Guide</strong> 101