8 Troubleshooting<br />
Reconfiguring Your Instrument<br />
7. Click the Communications button to reset the network IP address and assign a TCP<br />
port. If you have other equipment on your network and you need to move the <strong>ISQ</strong> to a<br />
different IP address, you can enter it here. For security purposes, you may also want to<br />
modify the TCP port. Consult your local IT Department for help.<br />
8. Click the Maintenance Intervals tab to set the number <strong>of</strong> days until you plan to perform<br />
maintenance on certain components <strong>of</strong> your GC/MS system. As a default, the foreline<br />
pump and turbo pump oil are automatically enabled. You can monitor the progress <strong>of</strong><br />
these settings in the Status pane <strong>of</strong> Xcalibur.<br />
9. Select the Foreline pump oil checkbox to enable the maintenance intervals. Then set the<br />
number <strong>of</strong> days in which you want to be reminded to check the oil. The manufacturer<br />
recommends changing the oil every 125 days. If you are using corrosive gases, such as<br />
ammonia, you should change the oil every 30 days.<br />
Thermo Scientific <strong>ISQ</strong> <strong>User</strong> <strong>Guide</strong> 101