j. Repeat the leak check until no leaks are indicated.<br />
16. Calibrate the carrier gas flow.<br />
1 Changing the Column<br />
Replacing the Factory-Installed Column<br />
a. Carefully push the capillary column end into the flow meter section <strong>of</strong> the column<br />
flowmeter connector, as shown below.<br />
Figure 4. Column Flowmeter Connector<br />
CAUTION - INSTRUMENT DAMAGE: Do not allow the column flowmeter connector to<br />
exceed 80 °C. Otherwise, it will melt and damage the instrument.<br />
b. Connect the flowmeter to the dedicated fitting on the column flow meter connector.<br />
c. Press Column Eval to open the following menu:<br />
COLUMN INFORMATION<br />
Right column<br />
Left column <<br />
Note The available options (Right column or Left column) will differ depending<br />
on the configuration <strong>of</strong> your GC. If you have a FOCUS GC, you do not need to<br />
select an option.<br />
Note Make sure the Vacuum Compensation parameter is set to OFF in the<br />
Carrier menu. Otherwise an accurate column evaluation cannot be performed.<br />
Thermo Scientific <strong>ISQ</strong> <strong>User</strong> <strong>Guide</strong> 7