Replacing the Factory-Installed Column<br />
To replace the factory-installed column in your <strong>ISQ</strong>:<br />
1 Changing the Column<br />
Replacing the Factory-Installed Column<br />
Note If you are running samples, stop the acquisition before powering <strong>of</strong>f the system.<br />
1. Cool down the GC oven and injector. See the GC documentation for information.<br />
2. Open the Xcalibur s<strong>of</strong>tware and click the Shutdown button in the lower left corner <strong>of</strong> the<br />
Status tab.<br />
3. Click the Yes button to continue the shutdown process. The heaters will be set to 120 °C<br />
and when they reach that temperature, the turbomolecular pump is powered <strong>of</strong>f. Then<br />
the foreline pump is powered <strong>of</strong>f.<br />
4. Reach around the right side <strong>of</strong> the instrument and push down on the power switch to<br />
power-<strong>of</strong>f the <strong>ISQ</strong>.<br />
Note If you are using hydrogen as a carrier gas, do NOT reach over the top <strong>of</strong> the<br />
instrument to power it <strong>of</strong>f. Instead, reach around the right side or go to the back <strong>of</strong> the<br />
instrument and flip down the power switch.<br />
Figure 2. Powering Off the <strong>ISQ</strong><br />
Vacuum Interlock Shield<br />
5. Open the front door <strong>of</strong> the instrument.<br />
Power Switch<br />
Vent Valve<br />
Knob<br />
6. Look behind the right side <strong>of</strong> the vacuum interlock shield and twist the vent valve knob<br />
one and a half times in a clockwise direction to open the vent.<br />
Thermo Scientific <strong>ISQ</strong> <strong>User</strong> <strong>Guide</strong> 3