14.12.2012 Views

ISQ User Guide - Write Frame of Mind

ISQ User Guide - Write Frame of Mind

ISQ User Guide - Write Frame of Mind

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

8 Troubleshooting<br />

Investigating Peak Issues<br />

Table 2. Troubleshooting Baseline Issues in Your Data, continued<br />

Behavior Cause Characteristic Remedy<br />

Noise<br />

General Column is contaminated Condition or replace the column.<br />

High-frequency<br />

Mass spectrum is<br />

electronically noisy<br />

Mass spectrum is<br />

chemically noisy<br />

Investigating Peak Issues<br />

Table 3. Troubleshooting Peak Issues in Your Data<br />

Detector gas is contaminated<br />

Check the purity <strong>of</strong> your gas and install appropriate<br />

filters.<br />

Electrometer is defective. Replace the electrometer.<br />

Temperature <strong>of</strong> the detector is<br />

higher than column’s maximum<br />

allowable temperature<br />

Reduce the temperature detector to the maximum<br />

allowable temperature <strong>of</strong> the column.<br />

Column fittings are loose Tighten connections accordingly.<br />

Electrometer is not properly<br />

zeroed or electrometer is<br />

damaged<br />

Column is dirty or MS is<br />

contaminated<br />

Behavior Cause Characteristic Remedy<br />

Broadening General<br />

Double Peaks General<br />

Re-zero the electrometer.<br />

Replace the electrometer, if necessary.<br />

Reset the electrometer by powering the system <strong>of</strong>f<br />

and on.<br />

Check for an air leak.<br />

Find the source <strong>of</strong> chemical noise and fix it.<br />

Check purity <strong>of</strong> carrier gas.<br />

Check the cleanliness <strong>of</strong> your inline carrier gas filter.<br />

Verify the GC liner, column and MS are free <strong>of</strong><br />

contamination.<br />

Verify that the transfer line temperature is not set too<br />

low.<br />

Column flow is too high<br />

Reduce the flow. Make sure vacuum compensation is<br />

turned on.<br />

Column flow is too low Increase the flow.<br />

Split flow is too low for split<br />

injection<br />

Increase the flow to 40-50 ml/min.<br />

Performance <strong>of</strong> the column has<br />

degraded<br />

Test the column at the optimum flow rate.<br />

Injector is dirty Clean or replace the liner.<br />

Temperature <strong>of</strong> the transfer line is<br />

too low<br />

Increase the temperature to 10 °C below the<br />

maximum allowable temperature <strong>of</strong> the column.<br />

Injection speed is too slow Inject more rapidly in a smooth motion.<br />

Autosampler injection speed or<br />

mode is incorrect<br />

Increase the speed and reduce the post injection<br />

dwell time.<br />

92 <strong>ISQ</strong> <strong>User</strong> <strong>Guide</strong> Thermo Scientific

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!