14.12.2012 Views

ISQ User Guide - Write Frame of Mind

ISQ User Guide - Write Frame of Mind

ISQ User Guide - Write Frame of Mind

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

Table 3. Troubleshooting Peak Issues in Your Data, continued<br />

Behavior Cause Characteristic Remedy<br />

Unresolved Peaks General<br />

GC oven is too cold<br />

Column has deteriorated<br />

Column is not far enough into the<br />

transfer line<br />

8 Troubleshooting<br />

Investigating Peak Issues<br />

The GC oven is not ramped to high enough<br />

temperature. Extend upper oven temperature.<br />

The GC column needs to be replaced or does not have<br />

the appropriate stationary phase for your application.<br />

Change GC column.<br />

The GC column does not extend into the <strong>ISQ</strong> source.<br />

Use the column measuring tool to confirm the column<br />

length. If the end <strong>of</strong> the column is inside the transfer<br />

line, an excessive amount <strong>of</strong> GC effluent will contact<br />

the inside wall.<br />

Ion source and/or transfer line are not hot enough.<br />

Increase the transfer line temperature. It should be at<br />

Transfer line is too cold<br />

least as hot as the highest GC oven temperature.<br />

Then try increasing the ion source temperature.<br />

Sample analyte is absorbing in the GC injector. Clean<br />

Injector is too active<br />

and deactivate the injection liner. You can also try<br />

liners made <strong>of</strong> different materials.<br />

Injection technique is not<br />

Choose a more appropriate injection technique.<br />

adequate<br />

Incorrect column flow rate Make sure the vacuum compensation is turned on.<br />

Thermo Scientific <strong>ISQ</strong> <strong>User</strong> <strong>Guide</strong> 95

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!