14.12.2012 Views

ISQ User Guide - Write Frame of Mind

ISQ User Guide - Write Frame of Mind

ISQ User Guide - Write Frame of Mind

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

3 Tuning the <strong>ISQ</strong><br />

Tuning the <strong>ISQ</strong> with Xcalibur<br />

6. Select the tune type you want to use:<br />

• Daily Tune Check—This tune is used to check how well your last tune is<br />

performing. It is the fastest tune type. The daily tune check performs a leak check,<br />

makes sure the mass calibration is okay, and sets the detector sensitivity to generate a<br />

m/z 69 ion with an intensity <strong>of</strong> 20,000,000 counts. If your SOP allows it, you can<br />

use this tune to rapidly verify that the previous lens tune is still generating good<br />

spectra.<br />

• Daily Tune—This tune is used to quickly tune the system. It performs a mass<br />

calibration and leak check, tunes the lenses and resolution, and sets the detector<br />

sensitivity to generate a m/z 69 ion with an intensity <strong>of</strong> 20,000,000 counts. You can<br />

perform a daily tune as frequently as your SOP requires. If the system is tuning to<br />

your satisfaction, then there is no need to perform the more time-consuming default<br />

or full tunes.<br />

• EI Default Tune—This tune creates a default tune file. It requires a clean<br />

instrument, starts with factory tune, and sets repeller to 0V and quadrupole voltage<br />

to a low value. This tune is used to generate a base for all the other tunes. As a result,<br />

you should only use this tune when the ion source is clean. The EI default tune will<br />

start with the tune file stored in the instrument at the factory and then perform a<br />

mass calibration and leak check, set the repeller to 0 V and tune the lenses. The<br />

quadrupole <strong>of</strong>fset voltage will be set to a low value to improve resolution, which is<br />

also tuned. The detector gain will be calibrated to generate 300,000 electrons for<br />

every ion that strikes the detector. This tune is used in cases where a dirty ion source<br />

has been replaced by a clean one or where the computer has been replaced.<br />

Additionally, this tune will generate spectra that are the closest in appearance to the<br />

factory tune.<br />

Note If you are starting with a clean source, this is the recommended tune to run.<br />

• EI Full Tune—This tune is used to completely retune the system. It takes the longest<br />

amount <strong>of</strong> time to run, but its advantage is that it re-optimizes nearly all the<br />

parameters affecting the signal. This type <strong>of</strong> tune will perform a mass calibration,<br />

tune the lenses and resolution, and perform a leak check. The detector gain will be<br />

calibrated to generate 300,000 electrons for every ion that strikes the detector. You<br />

should run an EI full tune when the daily tune or the daily tune check is not<br />

adequate, when the electron multiplier is getting old (tuning to high electron<br />

multiplier voltages), or the first tune after you replace the electron multiplier. Unless<br />

your SOP requires it, this is not the best tune to use on a daily basis because <strong>of</strong> the<br />

length <strong>of</strong> time it takes to run it.<br />

26 <strong>ISQ</strong> <strong>User</strong> <strong>Guide</strong> Thermo Scientific

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!