Thermocouple <strong>Measure</strong>ments<br />

Features and Functions 3<br />

The instrument supports the following thermocouple types: B, E, J, K, N,<br />

R, S, and T using ITS- 90 software conversions. The default is a J- Type<br />

thermocouple.<br />

Thermocouple measurements require a reference junction temperature.<br />

For the reference junction temperature, you can use an internal<br />

measurement on the module (34921A only), an external thermistor <br />

or RTD measurement, or a known fixed junction temperature.<br />

The internal reference junction source is valid only on channels 1<br />

through 40 on the 34921A with the 34921T terminal block installed.<br />

If you select an external reference, the instrument makes<br />

thermocouple measurements relative to a previously- stored RTD or<br />

thermistor measurement stored in a reference register. To store a<br />

reference temperature, first configure a multiplexer channel for an<br />

RTD or thermistor measurement. Then assign the measurement <br />

from that channel as the external reference. When you initiate a<br />

measurement on an external reference channel, the acquired<br />

temperature is stored in volatile memory in the reference register.<br />

Subsequent thermocouple measurements use the stored temperature<br />

as their reference. The temperature remains in memory until you<br />

measure a subsequent external reference value in the reference<br />

register or remove the mainframe power.<br />

If you select a fixed reference temperature, specify a value between <br />

- 20 °C and +80 °C (always specify the temperature in °C regardless<br />

of the temperature units currently selected).<br />

The accuracy of the measurement is highly dependent upon the<br />

thermocouple connections and the type of reference junction used.<br />

Use a fixed temperature reference for the highest accuracy<br />

measurements (you must maintain the known junction temperature).<br />

The internal isothermal block reference (34921A only) requires no<br />

external wiring but provides lower accuracy measurements than a fixed<br />

reference.<br />

The thermocouple check feature allows you to verify that your<br />

thermocouples are properly connected for measurements. If you enable<br />

this feature, the instrument measures the channel resistance after <br />

each thermocouple measurement to ensure a proper connection. If an<br />

open connection is detected (greater than 5 k on the 10 k range),<br />

the instrument reports an overload condition for that channel <br />

(or displays “OPEN T/C” on the front panel).<br />

<strong>Agilent</strong> <strong>34980A</strong> Mainframe User’s Guide 97