4 Introduction to the Plug-In Modules for the <strong>34980A</strong><br />

Electrical Operating Conditions<br />

WARNING<br />

Transients<br />

The 34921A, 34922A, 34923A, 34924A, 34925A, 34931A, 34932A, 34933A,<br />

34934A, 34937A, 34938A and 34939A modules are designed to safely<br />

withstand occasional transient overvoltages up to 1000 Vpeak. Typically,<br />

these transient overvoltages result from switching inductive loads or from<br />

nearby lightning strikes. The lightning- caused transient overvoltages that<br />

may occasionally occur on mains power outlets may be as high as 2500<br />

Vpeak.<br />

The 34941A, 34942A, 34945A, 34946A, 34947A, 34950A, 34951A, 34952A<br />

and 34959A modules are intended for only low- voltage applications, and<br />

should not be connected to circuits that may generate or conduct large<br />

transient voltages.<br />

WARNING<br />

High Energy Sources<br />

To avoid electric shock, turn off the <strong>34980A</strong> and disconnect or<br />

de-energize all field wiring to the modules and the Analog Bus<br />

connector before removing any module or slot cover.<br />

These modules are designed to handle inputs up to their rated currents<br />

or their rated powers, whichever is less. Under certain fault conditions,<br />

high energy sources could provide substantially more current or power<br />

than a module can handle. It is important to provide external current<br />

limiting, such as fuses, if the module inputs are connected to high- energy<br />

sources.<br />

CAUTION<br />

Do not connect any of the modules directly to a mains power<br />

outlet. If it is necessary to measure a mains voltage or any circuit<br />

where a large inductive load may be switched, you must add<br />

signal conditioning elements to reduce the potential transients<br />

before they reach the module or the Analog Buses.<br />

Install current limiting devices between high energy sources and<br />

the module inputs.<br />

170 <strong>Agilent</strong> <strong>34980A</strong> Mainframe User’s Guide