4 Introduction to the Plug-In Modules for the <strong>34980A</strong><br />

Module Considerations<br />

This section lists important items and actions that can affect the<br />

operation of your modules.<br />

General Considerations<br />

NOTE<br />

Environmental Operating Conditions<br />

These modules are designed to operate in a temperature range of 0 °C to<br />

+55 °C with non- condensing humidity. The maximum humidity is 80% at<br />

40 °C or higher. Do not use in locations where conductive dust or<br />

electrolytic salt dust may be present.<br />

These modules should be operated in an indoor environment where<br />

temperature and humidity are controlled. Condensation can pose a<br />

potential shock hazard. Condensation can occur when the modules are<br />

moved from a cold to a warm environment, or if the temperature and/or<br />

humidity of the environment changes quickly.<br />

The table on page 169 shows maximum voltage, current and power<br />

dissipation ratings for each module. If conditions change, ensure that<br />

condensation has evaporated and the instrument has thermally stabilized<br />

until pollution degree 1 conditions are restored before turning on power<br />

to the equipment.<br />

NOTE<br />

NOTE<br />

CAUTION<br />

To reduce wear on the internal DMM relays, wire like functions<br />

on adjacent channels.<br />

Pollution Degree 1: No pollution or only dry, non-conductive<br />

pollution occurs. The pollution has no influence (on insulation)<br />

(IEC 61010-1 2nd Edition).<br />

Pollution Degree 2: Normally only non-conductive pollution<br />

occurs. Occasionally, a temporary conductivity (leakage current<br />

between isolated conductors) caused by condensation can be<br />

expected (IEC 61010-1 2nd Edition).<br />

For proper module cooling, all unused slots must be covered.<br />