3 Features and Functions<br />

Remote Interface Operation:<br />

You can use the MEASure? command without specifying a to<br />

quickly take a stand- alone DMM reading. Note, however, that with the<br />

MEASure? command, most measurement parameters are set to their<br />

default values.<br />

To close the desired channel relays and Analog Bus relays, use the<br />

ROUTe:CLOSe command. The Analog Bus relays on the multiplexer and<br />

matrix modules are numbered s911, s912, s913, etc.<br />

To directly control all measurement parameters or triggering, use the<br />

CONFigure, SENSe, and TRIGger commands without specifying a<br />

parameter. To initiate the measurement, use the INITiate or<br />

READ? command without specifying a . Each time you initiate a<br />

new measurement, the instrument will clear the previous set of<br />

readings from memory.<br />

To stop a measurement in progress, use the ABORt command.<br />

To view the readings in memory, use the FETCh? command<br />

(the readings are not erased when you read them).<br />

Scanning Mode<br />

In the Scanning Mode, the <strong>34980A</strong> automatically controls a sequence<br />

of measurements using the internal DMM, possibly across multiple<br />

channels, and stores the results in memory. The <strong>34980A</strong> closes and <br />

opens the appropriate channel relays and Analog Bus relays required <br />

for the sequence. The following general rules apply to the Scanning Mode<br />

(for more information on using the Scanning Mode, see “Scanning” on<br />

page 108.)<br />

Any channel that can be “read” by the instrument can also be included<br />

in a scan. A scan can also include a read of a digital channel or a read<br />

of the totalizer count on the digital modules.<br />

Before you can initiate a scan, you must set up a scan list to include all<br />

desired multiplexer or digital channels. Channels which are not in the<br />

scan list are skipped during the scan.<br />

The Analog Bus relays are automatically opened and closed as required<br />

during the scan to connect to the internal DMM for the measurement.<br />

For example, all 2- wire measurements use the ABus1 (MEAS) relays<br />

for 4- wire measurements, the ABus2 (SENS) relays are used in addition<br />

to the ABus1 relays.<br />

Each time you initiate a new scan, the instrument will clear the<br />

previous set of readings from memory.<br />

82 <strong>Agilent</strong> <strong>34980A</strong> Mainframe User’s Guide