Analog Bus and Internal DMM Considerations<br />

This section provides important environmental and electrical<br />

considerations that can affect mainframe operation.<br />

Environmental Operating Conditions<br />

Features and Functions 3<br />

The <strong>34980A</strong> mainframe, including the optional internal DMM, is designed<br />

to operate in a temperature range of 0 °C to +55 °C with non- condensing<br />

humidity. The maximum humidity is 80% at 40 °C or higher. Do not use in<br />

locations where conductive dust or electrolytic salt dust may be present.<br />

The <strong>34980A</strong> should be operated in an indoor environment where<br />

temperature and humidity are controlled. Condensation can pose a<br />

potential shock hazard. Condensation can occur when the instrument is<br />

moved from a cold to a warm environment, or if the temperature and/or<br />

humidity of the environment changes quickly.<br />

When used in pollution degree 1 conditions, the maximum voltage rating<br />

for the Analog Buses is 300V. When used in pollution degree 2 conditions,<br />

the maximum voltage rating is 100V. If conditions change, ensure that<br />

condensation has evaporated and the instrument has thermally stabilized<br />

until pollution degree 1 conditions are restored before turning on power to<br />

the equipment.<br />

NOTE<br />

NOTE<br />

Pollution Degree 1: No pollution or only dry, non-conductive pollution<br />

occurs. The pollution has no influence (on insulation) (IEC 61010-1 <br />

2nd Edition).<br />

Pollution Degree 2: Normally only non-conductive pollution occurs.<br />

Occasionally, a temporary conductivity (leakage current between isolated<br />

conductors) caused by condensation can be expected (IEC 61010-1 <br />

2nd Edition).<br />

<strong>Agilent</strong> <strong>34980A</strong> Mainframe User’s Guide 79