4 Introduction to the Plug-In Modules for the <strong>34980A</strong><br />
Slot and Channel Addressing Scheme<br />
The eight module slots in the <strong>34980A</strong> are arranged as shown below. <br />
Slot Number Indicators<br />
The slot and channel addressing scheme for the <strong>34980A</strong> follows the form<br />
sccc where s is the mainframe slot number (1 through 8) and ccc is the<br />
three- digit channel number. Note that MUX channels numbers are<br />
derived differently from matrix modules, and channel numbers for matrix<br />
modules are derived differently between 1- wire and 2- wire configuration<br />
modes.<br />
Displayed Number... Means This... Determined by...<br />
1014 A MUX module is in slot 1, channel of<br />
interest is 14. This channel is labeled on<br />
the simplified schematics as 014 on<br />
Bank 1 of each MUX module.<br />
3921 A MUX or matrix module is in slot 3,<br />
channel of interest is 921 (Analog Bus<br />
relay on ABus1)<br />
5304 A 34931A, 34932A, 34933A (2-wire<br />
mode) matrix module is in slot 5,<br />
crosspoint is row 3, column 4.<br />
2437 A 34933A matrix module in 1-wire<br />
mode is in slot 2, matrix of interest is 4,<br />
crosspoint is row 3, column 7.<br />
MUX module channel numbers are<br />
determined by the numbers assigned to<br />
the switches on each bank. Channel<br />
numbers contain three digits.<br />
MUX and matrix channel numbers for the<br />
Analog Bus relays are determined by the<br />
number assigned to the relays.<br />
Matrix module (in 2-wire mode) channel<br />
numbers are derived from the crosspoint<br />
or intersection of rows and columns,<br />
columns having two digits).<br />
34933A matrix module (in 1-wire mode)<br />
channel numbers are derived from a<br />
specific matrix number and the crosspoint<br />
or intersection of rows and columns on<br />
that matrix.<br />
166 <strong>Agilent</strong> <strong>34980A</strong> Mainframe User’s Guide<br />
-