Agilent 34980A Multifunction Switch/Measure Unit *34980-90005*
Agilent 34980A Multifunction Switch/Measure Unit *34980-90005*
Agilent 34980A Multifunction Switch/Measure Unit *34980-90005*
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
Automatic Trigger Delays<br />
Features and Functions 3<br />
If you do not specify a trigger delay, the instrument selects a delay<br />
for you. The delay is determined by the function, range, integration time,<br />
and ac filter setting as shown below.<br />
DC Voltage, Thermocouple, DC Current (for all ranges):<br />
Integration Time Trigger Delay<br />
PLC > 1<br />
PLC 1<br />
Resistance, RTD, Thermistor (2- and 4-wire):<br />
Range<br />
100<br />
1 k<br />
10 k<br />
100 k<br />
1 M<br />
10 M<br />
100 M<br />
AC Voltage, AC Current (for all ranges):<br />
Frequency, Period:<br />
Digital Input, Totalize:<br />
2.0 ms<br />
1.0 ms<br />
Trigger Delay <br />
(for PLC > 1)<br />
2.0 ms<br />
2.0 ms<br />
2.0 ms<br />
25 ms<br />
30 ms<br />
200 ms<br />
200 ms<br />
Trigger Delay <br />
(for PLC 1)<br />
1.0 ms<br />
1.0 ms<br />
1.0 ms<br />
20 ms<br />
25 ms<br />
200 ms<br />
200 ms<br />
AC Filter Trigger Delay<br />
Slow (3 Hz)<br />
Medium (20 Hz)<br />
Fast (200 Hz)<br />
7.0 seconds<br />
1.0 second<br />
120 ms<br />
AC Filter Trigger Delay<br />
Slow (3 Hz)<br />
Medium (20 Hz)<br />
Fast (200 Hz)<br />
Trigger Delay<br />
0 seconds<br />
600 ms<br />
300 ms<br />
100 ms<br />
<strong>Agilent</strong> <strong>34980A</strong> Mainframe User’s Guide 91