3 Features and Functions<br />

At the end of the scan, the last channel that was scanned will be<br />

opened (as well as any Analog Bus relays used during the scan). <br />

Any channels that were opened during the scan will remain open at <br />

the completion of the scan.<br />

If you abort a scan that is running, the instrument will terminate any<br />

reading in progress (readings are not cleared from memory). If a scan<br />

is in progress when the command is received, the scan will not be<br />

completed and you cannot resume the scan from where it left off. <br />

Note that if you initiate a new scan, all readings are cleared from memory.<br />

You can use either the internal DMM or an external instrument to make<br />

measurements of your configured channels. However, the <strong>34980A</strong> allows<br />

only one scan list at a time; you cannot scan some channels using the<br />

internal DMM and others using an external instrument. Readings are<br />

stored in <strong>34980A</strong> memory only when the internal DMM is used.<br />

The Monitor mode is automatically enabled on all channels that are<br />

part of the active scan list (see “Monitor Mode” on page 127).<br />

The present scan list is stored in volatile memory and will be lost when<br />

power is turned off or after a Factory Reset (*RST command).<br />

Adding Channels to the Scan List<br />

Before you can initiate a scan, you must set up a scan list to include all<br />

desired multiplexer or digital channels. Channels which are not in the<br />

scan list are skipped during the scan. By default, the instrument scans the<br />

list of channels in ascending order from slot 1 through slot 8 (channels<br />

are reordered as needed).<br />

To Build a Scan List From the Front Panel<br />

To add the active channel to the scan list, press Channel (Configure).<br />

Then select the function, range, resolution, and other parameters for<br />

this channel. Then add the channel to the scan list by selecting:<br />

SCAN THIS CHANNEL? > YES<br />

To remove the active channel from the scan list, select:<br />

SCAN THIS CHANNEL? > NO<br />

To remove all channels from the scan list, select:<br />

Scan (Configure) > CLEAR SCAN LIST? > YES<br />

To initiate a scan and store all readings in memory, press<br />

Scan (<strong>Measure</strong>). Each time you initiate a new scan, the instrument<br />

clears all previously stored readings. If you have not defined a scan list,<br />

Scan (<strong>Measure</strong>) performs an internal DMM scan independent of any<br />

channels.<br />

To stop a scan in progress, press and hold Scan (<strong>Measure</strong>).<br />