2 Getting Started<br />
Menu Example 3: Configuring the DMM for a <strong>Measure</strong>ment<br />
This example uses the internal DMM for a measurement. It can be used<br />
with any of the multiplexer modules (34921A, 34922A, 34923A, 34924A,<br />
or 34925A). If you have disabled your internal DMM, don't have one<br />
installed in your <strong>34980A</strong>, or are not using a multiplexer module, skip<br />
this example.<br />
1 Make sure you have a multiplexer module installed in slot 1. Select<br />
any channel on this module (the channel number is shown in green<br />
channel field on display).<br />
2 Within the Configure key group, press DMM to set the measurement<br />
parameters for the internal DMM. Note that the DMM key lights.<br />
2<br />
3 Using the knob to navigate through the menu choices, select the AC<br />
VOLTS function.<br />
4 Press the lighted DMM key to accept the ac voltage function.<br />
5 Using the knob, select 10V, which indicates the range parameter for ac<br />
voltage.<br />
6 Press EXIT MENU to save your selections (ac voltage, 10 volt range)<br />
and immediately exit the menu. All unmodified parameters within the<br />
menu remain unchanged.<br />