Calibration Message<br />

Features and Functions 3<br />

The instrument allows you to store one message in calibration memory <br />

in the mainframe, a digital module, or the internal DMM (remote interface<br />

operation only). For example, you can store such information as the date<br />

when the last calibration was performed, the date when the next<br />

calibration is due, the instrument’s serial number, or even the name and<br />

phone number of the person to contact for a new calibration.<br />

You can record a calibration message only from the remote interface<br />

and only when the instrument is unsecured. You can read the<br />

calibration message whether the instrument is secured or unsecured.<br />

The calibration message may contain up to 40 characters. From the<br />

front panel, you can view 18 characters of the message at a time.<br />

Storing a calibration message will overwrite any message previously<br />

stored in memory.<br />

The calibration message is stored in non- volatile memory in the<br />

mainframe, a digital module, or the internal DMM, and does not change<br />

when power has been off, after a Factory Reset (*RST command), <br />

or after an Instrument Preset (SYSTem:PRESet command).<br />

Remote Interface Operation: The following example shows how to store a<br />

message in calibration memory on the module in slot 3.<br />

CAL:STRING "CAL: 21 NOV 2005",3<br />

<strong>Agilent</strong> <strong>34980A</strong> Mainframe User’s Guide 157