Agilent 34980A Multifunction Switch/Measure Unit *34980-90005*
Agilent 34980A Multifunction Switch/Measure Unit *34980-90005*
Agilent 34980A Multifunction Switch/Measure Unit *34980-90005*
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
Calibration Message<br />
Features and Functions 3<br />
The instrument allows you to store one message in calibration memory <br />
in the mainframe, a digital module, or the internal DMM (remote interface<br />
operation only). For example, you can store such information as the date<br />
when the last calibration was performed, the date when the next<br />
calibration is due, the instrument’s serial number, or even the name and<br />
phone number of the person to contact for a new calibration.<br />
You can record a calibration message only from the remote interface<br />
and only when the instrument is unsecured. You can read the<br />
calibration message whether the instrument is secured or unsecured.<br />
The calibration message may contain up to 40 characters. From the<br />
front panel, you can view 18 characters of the message at a time.<br />
Storing a calibration message will overwrite any message previously<br />
stored in memory.<br />
The calibration message is stored in non- volatile memory in the<br />
mainframe, a digital module, or the internal DMM, and does not change<br />
when power has been off, after a Factory Reset (*RST command), <br />
or after an Instrument Preset (SYSTem:PRESet command).<br />
Remote Interface Operation: The following example shows how to store a<br />
message in calibration memory on the module in slot 3.<br />
CAL:STRING "CAL: 21 NOV 2005",3<br />
<strong>Agilent</strong> <strong>34980A</strong> Mainframe User’s Guide 157