3 Features and Functions<br />

Auto-IP<br />

The Auto- IP standard automatically assigns an IP address to the <strong>34980A</strong><br />

when on a network that does not have DHCP servers.<br />

If you change the Auto- IP configuration, you must cycle power on the<br />

<strong>34980A</strong> to activate the new setting. <br />

Auto- IP allocates IP addresses from the link- local address range<br />

(169.254.xxx.xxx).<br />

From the factory, the Auto- IP setting is enabled.<br />

The Auto- IP setting is stored in non- volatile memory, and does not<br />

change when power has been off, after a Factory Reset (*RST<br />

command), or after an Instrument Preset (SYSTem:PRESet command).<br />

Front Panel Operation:<br />

Utility > REMOTE I/O > LAN > LAN SETTINGS > MODIFY > DHCP OFF > AUTO IP<br />

Remote Interface Operation:<br />

SYSTem:COMMunicate:LAN:AUTOip (OFF|ON}<br />

72 <strong>Agilent</strong> <strong>34980A</strong> Mainframe User’s Guide