Introduction to the <strong>34980A</strong> 1<br />

The figure below shows the external connections required to synchronize<br />

the scan sequence between the <strong>34980A</strong> and an external instrument. The<br />

<strong>34980A</strong> must notify the external instrument when a relay is closed and<br />

fully settled (including channel delay). The <strong>34980A</strong> outputs a Channel<br />

Closed pulse. In response, the external instrument must notify the<br />

<strong>34980A</strong> when it has finished its measurement and is ready to advance to<br />

the next channel in the scan list. For more information, see “Scanning<br />

With External Instruments” on page 129.<br />

ABus1 HI<br />

ABus2 HI<br />

ABus3 HI<br />

ABus4 HI<br />

Analog Bus Connector Ext Trig Connector<br />

9<br />

6<br />

5<br />

1<br />

ABus1 LO<br />

ABus2 LO<br />

ABus3 LO<br />

ABus4 LO<br />

Channel Advance<br />

(In)<br />

GND<br />

Channel Closed<br />

(Out)<br />

<strong>34980A</strong> Mainframe<br />

<strong>Agilent</strong> <strong>34980A</strong> Mainframe User’s Guide 15<br />

6<br />

9<br />

1<br />

5<br />

External<br />

DMM<br />

VM Complete Out Ext Trig In