Features and Functions 3<br />
To ensure you are getting the most accurate measurements possible, <br />
use care when setting the channel delay less than the default value<br />
(automatic). The default channel delay is designed to optimize<br />
parameters, such as settling time, for the most accurate measurements.<br />
The CONFigure and MEASure? commands set the channel delay to<br />
automatic. A Factory Reset (*RST command) also sets the channel delay<br />
to automatic.<br />
Front Panel Operation: Channel (Configure) > CHANNEL DELAY > TIME<br />
Once you have added the specified channel to the scan list, the channel<br />
delay choice will be visible in the menu.<br />
Interface Operation: The following command add a 2- second channel delay<br />
to the specified channels.<br />
ROUT:CHAN:DELAY 2,(@1003,1013)<br />
Automatic Channel Delays<br />
If you do not specify a channel delay, the instrument selects a delay <br />
for you. The delay is determined by the delay based on function, range,<br />
integration time, and ac filter setting. <br />
DC Voltage, Thermocouple, DC Current (for all ranges):<br />
Integration Time Channel Delay<br />
PLC > 1<br />
PLC 1<br />
Resistance, RTD, Thermistor (2- and 4-wire):<br />
Range<br />
100<br />
1 k<br />
10 k<br />
100 k<br />
1 M<br />
10 M<br />
100 M<br />
2.0 ms<br />
1.0 ms<br />
Channel Delay <br />
(for PLC > 1)<br />
2.0 ms<br />
2.0 ms<br />
2.0 ms<br />
25 ms<br />
30 ms<br />
200 ms<br />
200 ms<br />
Channel Delay <br />
(for PLC 1)<br />
1.0 ms<br />
1.0 ms<br />
1.0 ms<br />
20 ms<br />
25 ms<br />
200 ms<br />
200 ms<br />
<strong>Agilent</strong> <strong>34980A</strong> Mainframe User’s Guide 121