Voltage (DAC) Output<br />

Introduction to the <strong>34980A</strong> 1<br />

The 34951A module has four independent, isolated analog output<br />

channels that can output dc voltage up to 16V or dc current up to <br />

20 mA. Each DAC (Digital- to- Analog Converter) channel can be used<br />

as a programmable voltage source for analog input control of other<br />

devices. A simplified diagram is shown below.<br />

16 Bits<br />

16 Bits<br />

16 Bits<br />

16 Bits<br />

DAC 1<br />

DAC 2<br />

DAC 3<br />

DAC 4<br />

DAC 1H<br />

DAC 1L<br />

DAC 2H<br />

DAC 2L<br />

DAC 3H<br />

DAC 3L<br />

DAC 4H<br />

DAC 4L<br />

You can set the output voltage to any value between - 16 Vdc and +16<br />

Vdc, in 500 V steps on any or all four channels.<br />

You can set the output current to any value between - 20 mA and +20<br />

mA, in 630 nA steps on any or all four channels.<br />

<strong>Agilent</strong> <strong>34980A</strong> Mainframe User’s Guide 19