Sweep Count<br />

Trigger Sweep 1 Sweep 2 Sweep n Trigger<br />

Sample count for Scanning Mode<br />

Features and Functions 3<br />

For scanning, the specified sample count sets the number of readings<br />

per channel (same for all channels in the scan list). If no channels have<br />

been assigned to the scan list, the sample count sets the number of<br />

readings per trigger for the internal DMM.<br />

You can specify a sample count in conjunction with a trigger count and<br />

a sweep count. The three parameters operate independent of one<br />

another, and the total number of readings returned will be the product<br />

of the three parameters.<br />

You can store at least 500,000 readings in memory and all readings are<br />

automatically time stamped. If memory overflows, the new readings will<br />

overwrite the first (oldest) readings stored; the most recent readings are<br />

always preserved.<br />

The CONFigure and MEASure? commands automatically set the sample<br />

count to 1.<br />

The instrument sets the sample count to 1 after a Factory Reset <br />

(*RST command). An Instrument Preset (SYSTem:PRESet command) or<br />

Card Reset (SYSTem:CPON command) does not change the setting.<br />

Front Panel Operation: Scan (Configure) > SAMPLE COUNT<br />

Remote Interface Operation: SAMPle:COUNt<br />

<strong>Agilent</strong> <strong>34980A</strong> Mainframe User’s Guide 119<br />

. . .<br />

Ch 1 Ch 2 Ch 3 Ch 4 Ch 5 Ch 6<br />

Sample Count<br />

(1 to 500,000 samples)<br />