Agilent 34980A Multifunction Switch/Measure Unit *34980-90005*
Agilent 34980A Multifunction Switch/Measure Unit *34980-90005*
Agilent 34980A Multifunction Switch/Measure Unit *34980-90005*
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
Sweep Count<br />
Trigger Sweep 1 Sweep 2 Sweep n Trigger<br />
Sample count for Scanning Mode<br />
Features and Functions 3<br />
For scanning, the specified sample count sets the number of readings<br />
per channel (same for all channels in the scan list). If no channels have<br />
been assigned to the scan list, the sample count sets the number of<br />
readings per trigger for the internal DMM.<br />
You can specify a sample count in conjunction with a trigger count and<br />
a sweep count. The three parameters operate independent of one<br />
another, and the total number of readings returned will be the product<br />
of the three parameters.<br />
You can store at least 500,000 readings in memory and all readings are<br />
automatically time stamped. If memory overflows, the new readings will<br />
overwrite the first (oldest) readings stored; the most recent readings are<br />
always preserved.<br />
The CONFigure and MEASure? commands automatically set the sample<br />
count to 1.<br />
The instrument sets the sample count to 1 after a Factory Reset <br />
(*RST command). An Instrument Preset (SYSTem:PRESet command) or<br />
Card Reset (SYSTem:CPON command) does not change the setting.<br />
Front Panel Operation: Scan (Configure) > SAMPLE COUNT<br />
Remote Interface Operation: SAMPle:COUNt<br />
<strong>Agilent</strong> <strong>34980A</strong> Mainframe User’s Guide 119<br />
. . .<br />
Ch 1 Ch 2 Ch 3 Ch 4 Ch 5 Ch 6<br />
Sample Count<br />
(1 to 500,000 samples)<br />
t